A system for intravascular, radially orientation-independent electromagnetic flow- and diameter-sensing.

作者信息

Kolin A

出版信息

Pflugers Arch. 1977 Nov 25;372(1):109-11. doi: 10.1007/BF00582215.

Abstract

The difficulty of optimally orienting an intravascular loop probe of an extracorporeal field electromagnetic flow meter is circumvented by uniting 2 mutually perpendicular loop sensors into a single flow-diameter probe. When one of the loops is unfavorably oriented in the magnetic field, the orientations of the other loop is more favorable. The most unfavorable case is a 45 degrees angle between the magnetic field and the planes of the loops, when the signal drops to 70.7% of the optimal value. By taking the square root of the sum of the squares of the 2 loop transducer signals, one obtains an output for flow and diameter measurements which is independent of the probe orientation. This operation can be accomplished electronically.