[三维计算机断层扫描测量肝体积的验证]

[Validation of the measurement of hepatic volume by three-dimensional computed tomography].

作者信息

Evrard S, Moyses B, Ghnassia J P, Vix M, Mutter D, Methelin G, Marescaux J

机构信息

Clinique Chirurgicale A, Les Hôpitaux Universitaires de Strasbourg.

出版信息

Ann Chir. 1992;46(7):601-4.

PMID:1456690
Abstract

The aim of this work was to verify experimentally the reliability of 3D CT scan in measuring hepatic volumes. Eight livers were dissected from corpses. The referring liver volume was determined by measure of its hydric shift. The correlation coefficient R was 0.93 (p < 0.001) on 5 mm/5 mm 3D sections, 0.92 (p < 0.001) on 10 mm/10 mm 3D sections and 0.94 (p < 0.0006) on 5 mm/10 mm 3D sections. The paired t-test also showed a 5% risk correlation between the two measurements. Although in vivo imaging is slightly different from in vitro, 3D CT scan seems to be a reliable method to determine an hepatic volume without increased X-ray irradiation.

摘要

这项工作的目的是通过实验验证三维CT扫描测量肝脏体积的可靠性。从尸体上解剖出8个肝脏。参考肝脏体积通过测量其水移位来确定。在5毫米/5毫米的三维切片上,相关系数R为0.93(p<0.001),在10毫米/10毫米的三维切片上为0.92(p<0.001),在5毫米/10毫米的三维切片上为0.94(p<0.0006)。配对t检验也显示两次测量之间存在5%的风险相关性。尽管体内成像与体外成像略有不同,但三维CT扫描似乎是一种可靠的方法,可在不增加X射线辐射的情况下确定肝脏体积。