用于颞下颌关节运动分析的光学轴描记法与传统轴描记法的比较分析

A comparative analysis of optical and conventional axiography for the analysis of temporomandibular joint movements.

作者信息

Wagner Arne, Seemann Rudolf, Schicho Kurt, Ewers Rolf, Piehslinger Eva

机构信息

University Hospital of Cranio-Maxillofacial and Oral Surgery, Medical School, University of Austria, Vienna.

出版信息

J Prosthet Dent. 2003 Nov;90(5):503-9. doi: 10.1016/s0022-3913(03)00482-7.

Abstract

STATEMENT OF PROBLEM

Currently available systems for pantographic tracing are heavy, bulky, and can interfere with jaw movements.

PURPOSE

This study describes the development and clinical application of optoelectronic axiography designed to overcome system inherent problems of conventional bulky frame-based registration axiography. The purpose of this study is the comparison of the newly developed system and conventional axiography.

MATERIAL AND METHODS

Three-dimensional recordings of condylar pathways were acquired by means of infrared digitizers interfaced to newly developed software. Ten distinct curves in each of 10 subjects were recorded by synchronous optoelectronic axiography (100 tracings) and by conventional axiography (100 tracings). Usually, two 3-dimensional (3D) light weight sensors are provisionally fixed to the facial surface of a maxillary and mandibular incisor by means of a single orthodontic bracket. To allow for direct comparison of all 100 pairs of curves in this study, the 3D sensors of the optoelectronic system were attached to the bulky double face-bow system of the axiograph. The conformity of tracings (protrusion, opening/closing, mediotrusion, and laterotrusion) was evaluated by means of correlation analysis. Resulting axiographic recordings from both systems were evaluated by 3 experts (dentists, experienced in axiographic investigations, who were blind to the source of the data), focusing on standardized qualitative criteria of the recordings (homogeneity/smoothness, pathway-characteristics, excursion, and left/right-symmetry). After testing for normal distribution of the ratio scaled data (length of pathway, horizontal condylar inclination [HCI], Bennett angle) with the Kolmogoroff-Smirnov test (alpha=.01), axiographic curves were quantitatively compared by means of an intraclass correlation coefficient ([ICC] alpha =.01). The Wilcoxon test (alpha=.01) was used to evaluate equivalence of ordinally scaled values (homogeneity of tracings) and Cohen's Kappa was used to compare excursion and left/right symmetry.

RESULTS

High correspondence between curves recorded by conventional and optoelectronic axiography was observed. The mean differences of lengths between the protrusive, opening/closing, and mediotrusive pathways were 0.0 mm, 0.6 mm, and 0.1 mm, respectively. Pathways and values for HCI were found highly correlated (pathways: 95% CI of ICC 0.9776-0.9908; HCI: 95% CI of ICC 0.8641-0.9597). The 95% CIs for differences of pathways, HCI-value, and Bennett angle were -0.1mm/0.3mm, -3.4 degrees/1.9 degrees, and -2.8 degrees/4.8 degrees, respectively. Pathway characteristics also corresponded well (Cohen's Kappa: 0.73 for symmetric and 0.72 for asymmetric movements), 0.77 for left/right symmetry, whereas other characteristics showed less significant correlation (Cohen's Kappa of excursion: 0.21 for symmetric and 0.09 for asymmetric movements, homogeneity: 0.08 for symmetric and 0.15 for asymmetric movements).

CONCLUSION

Within the limitations of this study, optoelectronic axiography proved to be an applicable, promising technique, leading to diagnostic interpretations equivalent (with respect to the CIs) to conventional axiography.

摘要

问题陈述

目前可用的全景描记系统笨重、体积大,且会干扰下颌运动。

目的

本研究描述了为克服传统笨重的基于框架的配准轴描记法的系统固有问题而设计的光电轴描记法的开发及临床应用。本研究的目的是比较新开发的系统与传统轴描记法。

材料与方法

髁突运动轨迹的三维记录通过与新开发软件连接的红外数字化仪获取。10名受试者每人的10条不同曲线分别通过同步光电轴描记法(100次描记)和传统轴描记法(100次描记)进行记录。通常,两个三维(3D)轻质传感器通过单个正畸托槽临时固定在上颌和下颌切牙的面部表面。为了在本研究中直接比较所有100对曲线,将光电系统的3D传感器连接到轴描记仪的笨重双面弓系统上。通过相关分析评估描记(前伸、开闭、正中侧方运动和侧方运动)的一致性。由3位专家(对轴描记研究有经验且对数据来源不知情的牙医)对两个系统得到的轴描记记录进行评估,重点关注记录的标准化定性标准(均匀性/平滑度、轨迹特征、偏移和左右对称性)。在用柯尔莫哥洛夫-斯米尔诺夫检验(α = 0.01)检验比率标度数据(轨迹长度、水平髁突倾斜度[HCI]、贝内特角)的正态分布后,通过组内相关系数([ICC]α = 0.01)对轴描记曲线进行定量比较。采用威尔科克森检验(α = 0.01)评估顺序标度值的等效性(描记的均匀性),并使用科恩kappa系数比较偏移和左右对称性。

结果

观察到传统轴描记法和光电轴描记法记录的曲线之间具有高度一致性。前伸、开闭和正中侧方运动轨迹长度的平均差异分别为0.0毫米、0.6毫米和0.1毫米。发现轨迹和HCI值高度相关(轨迹:ICC的95%置信区间为0.9776 - 0.9908;HCI:ICC的95%置信区间为0.8641 - 0.9597)。轨迹、HCI值和贝内特角差异的95%置信区间分别为-0.1毫米/0.3毫米、-3.4度/1.9度和-2.8度/4.8度。轨迹特征也具有良好的一致性(科恩kappa系数:对称运动为0.73,不对称运动为0.72),左右对称性为0.77,而其他特征的相关性不太显著(偏移的科恩kappa系数:对称运动为0.21,不对称运动为0.09;均匀性:对称运动为0.08,不对称运动为0.15)。

结论

在本研究的局限性范围内,光电轴描记法被证明是一种适用的、有前景的技术,在置信区间方面能得出与传统轴描记法等效的诊断解释。