PURPOSE

To evaluate the viability of the PowerRefractor as a screening tool for examining refractive errors in large samples of children.

METHODS

The variability of the PowerRefractor was estimated using four patients. The refractive error was determined using cyclopentolate and tropicamide as cycloplegic agents and compared to that determined in a non-cycloplegic situation. In a second study, the data provided by the PowerRefractor were compared to results obtained by autorefractor or retinoscopy for 150 children aged from 6 months to 5 years.

RESULTS

Variability study. Statistical analysis showed a statistically significant difference between cycloplegic and non-cycloplegic refraction for spherical and cylindrical refractive errors (p<0.0001 in all cases). There was no significant difference between the measurements made using tropicamide and cyclopentolate (p=0.33 and p=0.18, respectively). Comparison study. In 142 of 150 patients the difference between data obtained by the PowerRefractor and an autorefractor was within 1 D (spherical equivalent). However, there was a considerable difference between the data generated by the two methods in the remaining eight patients (up to 16 D).

CONCLUSIONS

The PowerRefractor proved to be a reliable tool for estimating refractive errors in young children. The apparatus is easy to handle and the simultaneous examination of both eyes makes the PowerRefractor ideal for obtaining data on refractive errors in large samples.

摘要

目的

评估PowerRefractor作为筛查工具在检测大量儿童屈光不正方面的可行性。

方法

使用4名患者评估PowerRefractor的变异性。使用环喷托酯和托吡卡胺作为睫状肌麻痹剂确定屈光不正,并与非睫状肌麻痹情况下确定的结果进行比较。在第二项研究中,将PowerRefractor提供的数据与150名6个月至5岁儿童的自动验光仪或检影验光结果进行比较。

结果

变异性研究。统计分析表明,对于球镜和柱镜屈光不正,睫状肌麻痹验光和非睫状肌麻痹验光之间存在统计学显著差异(所有情况下p<0.0001)。使用托吡卡胺和环喷托酯进行的测量之间没有显著差异(分别为p=0.33和p=0.18)。比较研究。在150名患者中的142名中,PowerRefractor和自动验光仪获得的数据之间的差异在1 D(球镜等效值)以内。然而,在其余8名患者中,两种方法产生的数据存在相当大的差异(高达16 D)。

结论

PowerRefractor被证明是评估幼儿屈光不正的可靠工具。该设备易于操作,双眼同时检查使PowerRefractor非常适合获取大量样本的屈光不正数据。