Ono Masaki, Nakajima Toshiaki, Itoh Yuji, Shimada Kenji, Yamato Susumu
Analytical Laboratory, Self-Medication Laboratories, Taisho Pharmaceutical Co., Ltd., Yoshino-Cho 1-Chome, Kita-Ku, Saitama-Shi, Saitama 331-9530, Japan.
J Pharm Biomed Anal. 2003 Dec 4;33(5):1175-80. doi: 10.1016/s0731-7085(03)00427-8.
A flow injection system for myo-inositol determination in multivitamin pharmaceutical preparations using two enzyme reactors was developed. Myo-inositol was detected using a fluorophotometer, to measure the fluorescence of NADH produced from NAD+ by a myo-inositol dehydrogenase reactor (IDR) containing myo-inositol dehydrogenase immobilized on porous glass. Enhanced interference due to excess glucose included in a multivitamin pharmaceutical preparation as a sweetener was eliminated by a glucose eliminating reactor (GER) co-immobilized with three enzymes (glucose oxidase, mutarotase and catalase). The calibration coefficient for the standard curve was 0.9993 for myo-inositol detection in the range of 1-5 microg/ml. Myo-inositol was determined even in the presence of glucose concentrations of 140-420 microg/ml. The recovery of myo-inositol added to the multivitamin pharmaceutical preparation was 99.6% (n=9).