Mathieu H, Le Moigne F, Panteix G, Derache P, Jouzier E
Laboratoire Bioffice, Bordeaux.
Ann Biol Clin (Paris). 2003 Nov-Dec;61(6):667-72.
A potentiometric stripping analysis method for blood lead determination, using Radiometer Tracelab PSU 22, has been investigated. Hemoglobin precipitation by hydrochloric and perchloric acids mixture allows to obtain the same signal for the same lead concentrations in aqueous standards and various blood samples. Thus it is possible to use a time saving calibration procedure. Equilibration time increase improved method sensitivity and precision. The analysis time was 3 minutes per sample. The detection limit was 0.006 microM/L. The relative standard deviation was 3.6% for 1.351 microM/L and 6% for 0.203 microM/L. The results of 66 determinations obtained by this method and by graphite furnace atomic absorption spectrometry were compared (R=0.991). Potentiometric stripping analysis is less expensive and very convenient for clinical laboratories.
研究了一种使用Radiometer Tracelab PSU 22进行血铅测定的电位溶出分析方法。盐酸和高氯酸混合物沉淀血红蛋白后,对于水标准溶液和各种血液样本中相同的铅浓度可获得相同的信号。因此可以采用节省时间的校准程序。平衡时间的增加提高了方法的灵敏度和精密度。每个样品的分析时间为3分钟。检测限为0.006微摩尔/升。对于1.351微摩尔/升,相对标准偏差为3.6%;对于0.203微摩尔/升,相对标准偏差为6%。比较了用该方法和石墨炉原子吸收光谱法进行66次测定的结果(R = 0.991)。电位溶出分析成本较低,对临床实验室非常方便。