Parker A J, Stevens J C, Clegg R T
Department of Otolaryngology, Head and Neck Surgery, Royal Hallamshire Hospital, Sheffield.
J Laryngol Otol. 1992 Oct;106(10):896-9. doi: 10.1017/s0022215100121218.
We present a validated method of determining in vivo pressure, flow and voice parameters from patients using tracheo-oesophageal voice prostheses post laryngectomy. This apparatus was constructed simply using equipment which should be available in most large Otolaryngology Departments. There was good agreement between measurements made over two separate sessions of in vivo opening pressure, maximum flow through the device and maximum voice amplitude and these were significant. These measurements may be useful in monitoring patient progress and also in elucidating the mechanics of prosthesis failure.
我们提出了一种经过验证的方法,用于测定喉切除术后使用气管食管语音假体患者的体内压力、流量和语音参数。该装置仅使用大多数大型耳鼻喉科科室应具备的设备构建而成。在两次单独的体内开口压力、通过该装置的最大流量和最大语音振幅测量之间存在良好的一致性,且这些测量结果具有显著性。这些测量结果可能有助于监测患者的进展情况,也有助于阐明假体失效的机制。