Abstract

The introduction of the variable-pitch feature on pulse oximeters in 1983 by the Nellcor Corporation (Hayward, CA) allowed users to rapidly detect changes in oxygen saturation by listening for changes in the pitch of the tones emitted by the pulse oximeter. A few individuals have reported that they have been unable to detect a change in pitch when oxygen saturation changes. To these individuals, the variable-pitch feature of these pulse oximeters has not been beneficial. Using the pitches from one manufacturer of oximeters, we created a computer program to simulate the pitches that accompanied various oxygen saturations. The pitches were recorded onto a tape player and played for 75 volunteer subjects unfamiliar with the pitches of a variable-pitch pulse oximeter. Of our sample, 67% were able to detect a single change in pitch corresponding to a 1% fall in oxygen saturation, and 11% of the population could not detect a change in pitch until there was a change in pitch with every beat. We suggested four alternative designs that may prove beneficial to this group of individuals.

摘要

1983年,Nellcor公司(加利福尼亚州海沃德市)在脉搏血氧仪上引入了可变音高功能,用户可以通过聆听脉搏血氧仪发出的音调变化来快速检测血氧饱和度的变化。有少数人报告称,当血氧饱和度发生变化时,他们无法检测到音调的变化。对于这些人来说,这些脉搏血氧仪的可变音高功能并没有起到作用。我们使用某一制造商生产的血氧仪的音调,创建了一个计算机程序来模拟与各种血氧饱和度相对应的音调。这些音调被录制到磁带播放器上,并播放给75名不熟悉可变音高脉搏血氧仪音调的志愿者听。在我们的样本中,67%的人能够检测到与血氧饱和度下降1%相对应的音调单次变化,11%的人直到每次心跳音调都发生变化时才能检测到音调变化。我们提出了四种可能对这组人有益的替代设计。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验