Sharma A K, Supe S S, Narayanan V K, Subbarangaiah K
Kiwdai Memorial Institute of Oncology, Bangalore, India.
Strahlenther Onkol. 1992 Jul;168(7):402-5.
Electron beams from an accelerator appear to originate from a point away from the vacuum window designated as virtual source. Virtual source to isocenter distances are determined experimentally by measurements in a polystyrene phantom with inverse square law method. Corrections required for percentage depth dose are also determined. Results of our study are compared with that of other investigators.