Time-resolved fluorescence measurements using microlens array and area imaging devices.

作者信息

Merk Susanne, Lietz Achim, Kroner Margareta, Valler Martin, Heilker Ralf

机构信息

Boehringer Ingelheim Pharma GmbH & Co. KG, Department of Integrated Lead Discovery, Birkendorfer Strasse 65, D-88397 Biberach an der Riss, Germany.

出版信息

Comb Chem High Throughput Screen. 2004 Feb;7(1):45-54. doi: 10.2174/138620704772884814.

Abstract

Time-resolved fluorescence (TRF) assay formats are frequently used technologies in high-throughput screening. In this article, we have characterised the novel Plate::Vision(2) 96-microlens array reader (Carl Zeiss Jena GmbH, Germany) and compared it to the novel LEADseeker Generation IV multimodality imaging system (LEADseeker Gen IV; Amersham Biosciences UK Ltd., UK) for applications in the TRF mode. In europium measurements using the TRF mode, the Plate::Vision displayed a limit of detection for europium of approximately 3 pM, which was comparable to two established TRF readers, the Discovery and the Victor V (both PerkinElmer Life Sciences Inc., USA). The LEADseeker's limit of detection only extended down to europium concentrations of approximately 10 pM in these experiments. For TRF resonance energy transfer (TR-FRET) experiments, a europium-biotin (Eu-biotin) conjugate was titrated with a streptavidin-allophycocyanin (SA-APC) conjugate. The Plate::Vision produced Z' values larger than 0.5 for the acceptor fluorophor emission with concentrations of Eu-biotin as low as 3 nM combined with 175 pM SA-APC. To achieve Z' values of at least 0.5 with the LEADseeker, concentrations of 10 nM Eu-biotin combined with SA-APC of at least 0.8 nM were required. In a drug screening application using TR-FRET, the energy transfer from a europium-labelled protein X (Eu-protein X) to a complex of biotinylated peptide Y with SA-APC was measured. Using the Plate::Vision, a Z' factor larger than 0.5 for the acceptor fluorophor emission was only obtained for a Eu-protein X concentration of at least 10 nM in combination with biotinylated peptide Y/SA-APC at saturating concentrations. Both the Plate::Vision and the LEADseeker show good quality results for applications in the TRF mode and enable an increased throughput based on their shortened measurement time in comparison to classic photomultiplier tube-based readers.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验