Broadband and band-selective IMPRESS-gHMBC: compensation of refocusing inefficiency with synchronized inversion sweep.

作者信息

Crouch Ronald, Boyer Robert D, Johnson Ross, Krishnamurthy Krish

机构信息

Applications Laboratory, Varian, Inc., Palo Alto, California 94303, USA.

出版信息

Magn Reson Chem. 2004 Mar;42(3):301-7. doi: 10.1002/mrc.1316.

Abstract

Compensation of refocusing inefficiency in a gHMBC experiment by replacing the rectangular pi pulse with a pair of adiabatic pulses with synchronized inversion sweep (CRISIS) significantly improves the performance of the gHMBC experiment. The CRISIS-gHMBC experiment retains the pure absorptive shapes in F1 and hence results in better lineshape and higher resolution than the current versions of magnitude mode gHMBC spectra. When used as a broadband experiment, CRISIS-gHMBC, owing to better refocusing efficiency of the adiabatic pulse pairs, gives improved performance across the 13C spectral width. Moreover, it is shown that CRISIS-gHMBC is a robust and improved alternative and when used along with the IMPRESS (Improved Resolution using Symmetrically Shifted pulses) technique further increases the sensitivity and resolution without additional experimental time. The IMPRESS-CRISIS combination is demonstrated for broadband gHMBC and band-selective gHMBC experiments. The ICbs-gHMBC [IMPRESS-CRISIS-band-selective gHMBC] experiment is an attractive and better alternative to individual band-selective gHMBC.