Kawasaki Y, Inada S
Department of Veterinary Medicine, Faculty of Agriculture, Kagoshima University, Japan.
Am J Vet Res. 1992 Jul;53(7):1096-100.
Effects of analog filter frequency on brain stem auditory-evoked potentials (BAEP) were investigated in 7 non-sedated dogs. The BAEP were recorded successively at various low-pass (LP) and high-pass (HP) filter frequency settings. The analog filters had a rolloff of 6 dB/octave. Decrease of LP filter frequency from 30 kHz to 100 Hz caused prolongation of the peak latency and reduction of the peak-to-peak (from a positive peak to the following trough) and absolute (from a positive peak to the baseline) amplitudes for all peaks, except the peak latency for P5 and the absolute amplitude for P4. Changes in these variables were statistically significant (P less than 0.05) at different cutoff frequencies specific for the individual peaks. The interpeak latency between P1 and P4, and P4/P1 peak-to-peak amplitude ratio were not changed significantly. At the lowest LP filter frequency of 100 Hz, positive peaks (fast waves) seemed to be superimposed on a slow positive wave (slow wave). In contrast, increase of HP filter frequency from 0.53 to 160 Hz did not result in significant changes for any peaks, except for reduction in the absolute amplitude of P4. The various effects of LP filter frequency and negligible effects of HP filter frequency on individual peaks may be attributable to their frequency composition and/or elimination of the slow wave at higher HP filter frequency settings. On the basis of our results, LP filter setting of 3 kHz and HP filter setting of less than or equal to 53 Hz are recommended for recording of BAEP in dogs.(ABSTRACT TRUNCATED AT 250 WORDS)
在7只未镇静的犬中研究了模拟滤波器频率对脑干听觉诱发电位(BAEP)的影响。在各种低通(LP)和高通(HP)滤波器频率设置下连续记录BAEP。模拟滤波器的滚降为6 dB/倍频程。将LP滤波器频率从30 kHz降低到100 Hz会导致所有波峰的峰潜伏期延长,峰峰值(从正峰到随后的波谷)和绝对幅度(从正峰到基线)降低,但P5的峰潜伏期和P4的绝对幅度除外。这些变量的变化在各个波峰特定的截止频率处具有统计学意义(P小于0.05)。P1和P4之间的峰间潜伏期以及P4/P1峰峰值幅度比没有明显变化。在最低的LP滤波器频率100 Hz时,正峰(快波)似乎叠加在一个慢正波(慢波)上。相比之下,将HP滤波器频率从0.53 Hz增加到160 Hz,除了P4的绝对幅度降低外,对任何波峰都没有导致显著变化。LP滤波器频率的各种影响以及HP滤波器频率对各个波峰的可忽略不计的影响可能归因于它们的频率组成和/或在较高HP滤波器频率设置下慢波的消除。根据我们 的结果,建议在犬中记录BAEP时LP滤波器设置为3 kHz,HP滤波器设置小于或等于53 Hz。(摘要截断为250字)