Suppr超能文献

一种用于1-20keV能量范围内计量学研究的可调谐单色X射线源:在衰减系数测量中的应用。

A tunable monochromatic X-ray source for metrological studies in the 1-20 keV energy range: application to the measurement of attenuation coefficients.

作者信息

Lépy Marie-Christine, Ferreux Laurent, Plagnard Johann

机构信息

BNM/Laboratoire National Henri Becquerel, CEA Saclay, Gif-Sur-Yvette Cedex F-91191, France.

出版信息

Appl Radiat Isot. 2004 Feb-Apr;60(2-4):159-65. doi: 10.1016/j.apradiso.2003.11.010.

Abstract

A tunable monochromatic X-ray source operating in the 1-20 keV energy range is described. An X-ray tube provides initial photons. A dispersive crystal performs the energy selection, according to Bragg's law. An X-ray detector is connected to the monochromator fixed exit. This setup can be used for metrological studies. A first application consists in measuring attenuation coefficients in the 4-10 keV energy range. Results for aluminum and copper are given, with average relative uncertainties (1sigma) of 1% and 3% respectively.

摘要

描述了一种在1 - 20 keV能量范围内工作的可调谐单色X射线源。一个X射线管提供初始光子。一块色散晶体根据布拉格定律进行能量选择。一个X射线探测器连接到单色仪的固定出口。该装置可用于计量学研究。第一个应用是测量4 - 10 keV能量范围内的衰减系数。给出了铝和铜的测量结果,平均相对不确定度(1σ)分别为1%和3%。

相似文献

2
Close-geometry efficiency calibration of p-type HPGe detectors with a Cs-134 point source.
Appl Radiat Isot. 2006 Oct-Nov;64(10-11):1303-6. doi: 10.1016/j.apradiso.2006.02.050. Epub 2006 Mar 23.
4
A prototype of a quasi-monochromatic system for mammography applications.
Phys Med Biol. 2005 May 21;50(10):2225-40. doi: 10.1088/0031-9155/50/10/003. Epub 2005 Apr 27.
5
Developing 152Eu into a standard for detector efficiency calibration.
Appl Radiat Isot. 2004 Feb-Apr;60(2-4):283-7. doi: 10.1016/j.apradiso.2003.11.029.
6
Measurement of energy and direction distribution of neutron and photon fluences in workplace fields.
Radiat Prot Dosimetry. 2004;110(1-4):237-41. doi: 10.1093/rpd/nch179.
7
Construction of 144, 565 keV and 5.0 MeV monoenergetic neutron calibration fields at JAERI.
Radiat Prot Dosimetry. 2004;110(1-4):85-9. doi: 10.1093/rpd/nch197.
8
Development of a two-dimensional virtual-pixel X-ray imaging detector for time-resolved structure research.
J Synchrotron Radiat. 2004 Mar 1;11(Pt 2):177-86. doi: 10.1107/S0909049503028863. Epub 2004 Feb 12.
9
X reference radiation qualities produced with tube voltages above 300 kV for the calibration and testing of dosemeters.
Radiat Prot Dosimetry. 2007;123(2):137-42. doi: 10.1093/rpd/ncl107. Epub 2006 Sep 23.
10
Neutron calibration facilities.
Radiat Prot Dosimetry. 2004;110(1-4):33-42. doi: 10.1093/rpd/nch215.