Dolanmaz Dogan, Uckan Sina, Isik Kubilay, Saglam Haci
Department of Oral and Maxillofacial Surgery, Faculty of Dentistry, Selcuk University, Postal Code 42079, Konya, Turkey.
Br J Oral Maxillofac Surg. 2004 Apr;42(2):127-32. doi: 10.1016/S0266-4356(03)00234-1.
Six unembalmed adult sheep mandibles were stripped of all soft tissues and sectioned at the midline. Each side had a sagittal split ramus osteotomy (SSRO) and was advanced 5 mm. Six of the hemimandibles were fixed with four-hole extended titanium miniplates and titanium screws, and the other six were fixed with four-hole extended absorbable plates and absorbable screws. All specimens were mounted in a servohydraulic testing unit, and a range of forces (0-140 N) was applied. Displacement of each proximal segment was recorded at 10 N increments from 0 to 140 N. Values for the two groups were compared using the Mann-Whitney U-test, and significant differences in displacement were seen only at loads between 10 and 50 N. The results indicate that when absorbable miniplates are used intermaxillary fixation may be necessary to stabilise the bony fragments in the early postoperative period.
六具未防腐处理的成年绵羊下颌骨被剥离所有软组织,并在中线处切开。每侧进行矢状劈开下颌支截骨术(SSRO),并向前推进5毫米。其中六个半下颌骨用四孔加长钛微型钢板和钛螺钉固定,另外六个用四孔加长可吸收板和可吸收螺钉固定。所有标本安装在伺服液压测试装置中,施加一系列力(0 - 140牛)。从0到140牛,以10牛的增量记录每个近端节段的位移。使用曼-惠特尼U检验比较两组的值,仅在10至50牛的载荷下观察到位移有显著差异。结果表明,当使用可吸收微型钢板时,术后早期可能需要颌间固定来稳定骨碎片。