使用Coapsys装置非体外循环治疗二尖瓣反流后二尖瓣环和左心室尺寸及左心室压力-容积关系的变化。

Changes in mitral annular and left ventricular dimensions and left ventricular pressure-volume relations after off-pump treatment of mitral regurgitation with the Coapsys device.

作者信息

Fukamachi Kiyotaka, Popović Zoran B, Inoue Masahiro, Doi Kazuyoshi, Schenk Soren, Ootaki Yoshio, Kopcak Michael W, McCarthy Patrick M

机构信息

Department of Biomedical Engineering/ND 20, Lerner Research Institute, The Cleveland Clinic Foundation, 9500 Euclid Avenue, Cleveland, OH 44195, USA.

出版信息

Eur J Cardiothorac Surg. 2004 Mar;25(3):352-7. doi: 10.1016/j.ejcts.2003.12.007.

Abstract

OBJECTIVE

The objective of this study was to evaluate the changes in mitral annular and left ventricular dimensions and left ventricular pressure-volume relations produced by the Myocor Coapsys device that has been developed to treat functional mitral regurgitation (MR) off-pump.

METHODS

The Coapsys device, which consists of anterior and posterior epicardial pads connected by a sub-valvular chord, was implanted in seven dogs with functional MR resulting from pacing induced cardiomyopathy. The Coapsys device was then sized by drawing the posterior leaflet and annulus toward the anterior leaflet. During sizing, MR grade was assessed using color flow Doppler echocardiography. Final device size was selected when MR was eliminated or minimized. Following implantation, heart failure was maintained by continued pacing for a period of 8 weeks. Mitral annular and left ventricular dimensions and left ventricular pressure-volume relations were evaluated by two-dimensional echocardiography and a conductance catheter, respectively, at pre-sizing, post-sizing, and after 8 weeks.

RESULTS

All implants were performed on beating hearts without cardiopulmonary bypass. Mean MR grade was reduced from 2.9+/-0.7 at pre-sizing to 0.7+/-0.8 at post-sizing (P<0.001), and was maintained at 0.8+/-0.8 after 8 weeks (P<0.01). The septal-lateral dimensions were significantly reduced at both mitral annular level [2.4+/-0.2 cm at pre-sizing, 1.5+/-0.3 cm at post-sizing (P<0.001) and 1.8+/-0.3 cm after 8 weeks (P<0.05)] and mid-papillary level [4.1+/-0.4 cm at pre-sizing, 2.4+/-0.2 cm at post-sizing (P<0.001) and 3.3+/-0.4 cm after 8 weeks (P<0.001)]. The end-systolic pressure-volume relation shifted leftward at post-sizing with a significantly steeper slope (P=0.03). There was a significant (P=0.03) leftward shift of the end-diastolic pressure-volume relation at post-sizing. After 8 weeks, these changes in pressure-volume relations tended to return to pre-sizing relations.

CONCLUSIONS

The Coapsys device significantly reduced MR by treating both the mitral annular dilatation and the papillary muscle displacement. Despite these significant dimensional changes, the Coapsys device did not negatively affect the left ventricular pressure-volume relations.

摘要

目的

本研究的目的是评估Myocor Coapsys装置所产生的二尖瓣环和左心室尺寸变化以及左心室压力-容积关系,该装置已被开发用于非体外循环下治疗功能性二尖瓣反流(MR)。

方法

将由经瓣下腱索连接的前、后心外膜垫组成的Coapsys装置植入7只因起搏诱导性心肌病导致功能性MR的犬体内。然后通过将后叶和瓣环拉向前叶来确定Coapsys装置的尺寸。在确定尺寸过程中,使用彩色多普勒超声心动图评估MR分级。当MR消除或减至最小时选择最终装置尺寸。植入后,通过持续起搏维持心力衰竭8周。分别在确定尺寸前、确定尺寸后和8周后,通过二维超声心动图和电导导管评估二尖瓣环和左心室尺寸以及左心室压力-容积关系。

结果

所有植入均在跳动的心脏上进行,无需体外循环。平均MR分级从确定尺寸前的2.9±0.7降至确定尺寸后的0.7±0.8(P<0.001),并在8周后维持在0.8±0.8(P<0.01)。二尖瓣环水平[确定尺寸前为2.4±0.2 cm,确定尺寸后为1.5±0.3 cm(P<0.001),8周后为1.8±0.3 cm(P<0.05)]和乳头肌中部水平[确定尺寸前为4.1±0.4 cm,确定尺寸后为2.4±0.2 cm(P<0.001),8周后为3.3±0.4 cm(P<0.001)]的室间隔-侧壁尺寸均显著减小。确定尺寸后,收缩末期压力-容积关系向左移位,斜率显著变陡(P=0.03)。确定尺寸后,舒张末期压力-容积关系显著向左移位(P=0.03)。8周后,这些压力-容积关系的变化趋于恢复到确定尺寸前的关系。

结论

Coapsys装置通过治疗二尖瓣环扩张和乳头肌移位显著降低了MR。尽管存在这些显著的尺寸变化,但Coapsys装置并未对左心室压力-容积关系产生负面影响。