Betsch H M, Aschoff A
Neurochirurgische Universitätsklinik Heidelberg.
Anasthesiol Intensivmed Notfallmed Schmerzther. 1992 Feb;27(1):51-2. doi: 10.1055/s-2007-1000251.
Artifacts were observed when using epidural pressure monitoring systems (Gaeltec ICT/B Titan). These were produced by radio waves from our beeper system. A non-existing electromagnetic shield of the Gaeltec-measuring amplifier SN 10/11 could be found as the cause. The artifacts can lead to therapeutical errors.