[盖尔特克颅内压监测仪中因个人传呼机系统无线电波导致的测量伪影]

[Measurement artifacts in Gaeltec intracranial pressure monitors due to radio waves from personal beeper systems].

作者信息

Betsch H M, Aschoff A

机构信息

Neurochirurgische Universitätsklinik Heidelberg.

出版信息

Anasthesiol Intensivmed Notfallmed Schmerzther. 1992 Feb;27(1):51-2. doi: 10.1055/s-2007-1000251.

Abstract

Artifacts were observed when using epidural pressure monitoring systems (Gaeltec ICT/B Titan). These were produced by radio waves from our beeper system. A non-existing electromagnetic shield of the Gaeltec-measuring amplifier SN 10/11 could be found as the cause. The artifacts can lead to therapeutical errors.

摘要

在使用硬膜外压力监测系统(Gaeltec ICT/B Titan)时观察到了伪迹。这些伪迹是由我们传呼机系统发出的无线电波产生的。可以发现Gaeltec测量放大器SN 10/11不存在电磁屏蔽是其原因。这些伪迹可能导致治疗错误。