Prycek Jifí, Ciganek Miroslav, Simek Zdenĕk
Faculty of Chemistry, Institute of Chemistry and Technology of Environmental Protection, Brno University of Technology, 118 Purkynova, 612 00 Brno, Czech Republic.
J Chromatogr A. 2004 Mar 19;1030(1-2):103-7. doi: 10.1016/j.chroma.2003.12.037.
The possibilities of utilising pressurised liquid extraction for five nitro-polycyclic aromatic hydrocarbons from an inert matrix are shown. Different extraction temperatures and pressures were tested. The highest recoveries were obtained at extraction pressure 14 MPa and temperature 100 C. Separation of non-polar, aromatic and polar fractions by the silica gel column chromatography is shown. n-Hexane, cyclohexane and dichloromethane as a solvent were tested. The best separations of monitored fractions were obtained, when extract was dissolved in cyclohexane. Non-polar and aromatic fractions eluted together when the extract was dissolved in dichloromethane.
展示了利用加压液体萃取从惰性基质中提取五种硝基多环芳烃的可能性。测试了不同的萃取温度和压力。在萃取压力14MPa和温度100℃时获得了最高回收率。展示了通过硅胶柱色谱法分离非极性、芳香族和极性馏分的过程。测试了正己烷、环己烷和二氯甲烷作为溶剂的情况。当提取物溶解在环己烷中时,对监测馏分的分离效果最佳。当提取物溶解在二氯甲烷中时,非极性和芳香族馏分一起洗脱。