Suppr超能文献

激光烧蚀电感耦合等离子体原子发射光谱法(LA-ICP-AES)分析硅酸盐岩石和石灰岩时内标物的选择

Selection of internal standards for analysis of silicate rocks and limestones by laser ablation inductively coupled plasma atomic emission spectrometry (LA-ICP-AES).

作者信息

Kanicky V, Mermet J M

机构信息

Department of Anal. Chemistry, Masaryk University, Kotlarska 2, CZ-61137, Brno, Czech Republic.

出版信息

Anal Bioanal Chem. 1996 Jul;355(7-8):887-8. doi: 10.1007/s0021663550887.

Abstract

Included (Si, Ca) and added (Li, B) internal standards (IS) have been used comparatively in LA-ICP-AES of silicates and limestones to improve both precision and accuracy. A Q-switched Nd:YAG laser (355 nm, 10 Hz, 10 mJ per shot) has been applied for ablation. Samples have been prepared by fusion with Li(2)B(4)O(7) and measured using a Perkin-Elmer Optima 3000 ICP system. Both types of IS have given a relative standard deviation (RSD) less than 2-1%, which improved the repeatability by a factor of 2-10 and calibration graphs have been linear over the whole concentration range.

摘要

在硅酸盐和石灰石的激光烧蚀电感耦合等离子体发射光谱分析(LA-ICP-AES)中,已对包含的(硅、钙)和添加的(锂、硼)内标(IS)进行了比较使用,以提高精密度和准确度。已应用调Q钕:钇铝石榴石激光(355纳米,10赫兹,每次脉冲10毫焦)进行烧蚀。样品通过与Li₂B₄O₇熔融制备,并使用珀金埃尔默Optima 3000 ICP系统进行测量。两种类型的内标均给出了小于2%至1%的相对标准偏差(RSD),将重复性提高了2至10倍,并且校准曲线在整个浓度范围内呈线性。