Technical failure of biliary metal stent deployment in a series of 116 applications.

作者信息

Bethge N, Wagner H J, Knyrim K, Zimmermann H B, Starck E, Pausch J, Vakil N

机构信息

Department of Internal Medicine IV, Hospital Neukölln, Berlin, Germany.

出版信息

Endoscopy. 1992 Jun;24(5):395-400. doi: 10.1055/s-2007-1010506.

Abstract

Biliary metal stents are thought to offer improved long-term palliation of malignant biliary obstruction due to a lower incidence of migration and clogging. Placement of these stents is technically more complicated than that of plastic endoprostheses and requires two experienced physicians. We report the incidence and reasons for apparent malfunction of expandable metal stent deployment (Wallstents and Strecker stents). In 116 applications of 82 Wallstents (endoscopic approach: n = 33, transhepatic approach: n = 49), we observed 19 cases of stent malfunction due to technical problems of stent delivery. In 13 cases (15.8%), the restraining membrane of the Wallstent could not be retracted sufficiently to deliver the stent. There were 6 (17.6%) failures in 34 cases of Strecker stent deployment. In 3 cases, we noted difficult balloon removal, including avulsion of the balloon catheter shaft within the endoscope during attempted balloon removal in one case. In one case, the Strecker stent could only be released partially, requiring subsequent endoscopic extraction. In two patients, only partial expansion of one end of the Strecker stent could be achieved. Given the significant malfunction rate of expandable metal stents during stent delivery, further improvements in the delivery system of the metal stents are required.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验