Klimowicz Adam, Bielecka-Grzela Stanislawa, Groth Lotte, Benfeldt Eva
Dermatopharmacotherapy Division, Department of Dermatology, Pomeranian Medical University, 72 Powstancow Wlkp. Str., PL-70-111 Szczecin, Poland.
Skin Res Technol. 2004 May;10(2):104-8. doi: 10.1111/j.1600-0846.2004.00058.x.
BACKGROUND/PURPOSE: For microdialysis studies in the skin, laboratory-made linear probes are often used. Probes can be assembled with or without a stainless-steel guide wire in the lumen of the dialysis fibre. From a theoretical viewpoint, this will alter the performance of the probe. The aim of this study was to compare the in vitro microdialysis recovery using probes with and without an internal guide wire, employing two different model drugs and a range of perfusate flow rates.
In vitro microdialysis recovery experiments were performed with and without internal guide wires at perfusion flow rates from 1 to 15 microL/min, using either salicylic acid or metronidazole as test compounds at 5 mg/L. Phosphate buffer (pH 7.4) was used as perfusate. Dialysates were analysed by a validated HPLC assay with spectrophotometric detection.
For both test compounds, the mean microdialysis recovery did not differ significantly between microdialysis probes with and without an internal guide wire at lower flow rates. At very high flow rates (10-15 microL/min), a significant difference in recovery could be seen for probes with internal guide wires when compared with probes without.
The use of a stabilizing internal guide wire in linear microdialysis probes did not influence the in vitro recovery of microdialysis at volumetric flow rates below 10 microL/min. These are the flow rates most frequently used for in vivo microdialysis studies in the skin.
背景/目的:在皮肤微透析研究中,常使用实验室自制的线性探针。透析纤维管腔内的探针可以在有或没有不锈钢导丝的情况下组装。从理论角度来看,这会改变探针的性能。本研究的目的是使用两种不同的模型药物和一系列灌注液流速,比较有无内部导丝的探针的体外微透析回收率。
以水杨酸或甲硝唑作为5mg/L的测试化合物,在1至15μL/min的灌注流速下,分别使用有和没有内部导丝的探针进行体外微透析回收率实验。磷酸盐缓冲液(pH 7.4)用作灌注液。透析液通过经过验证的带有分光光度检测的高效液相色谱法进行分析。
对于两种测试化合物,在较低流速下,有和没有内部导丝的微透析探针的平均微透析回收率没有显著差异。在非常高的流速(10 - 15μL/min)下,与没有内部导丝的探针相比,有内部导丝的探针在回收率上存在显著差异。
在低于10μL/min的体积流速下,线性微透析探针中使用稳定的内部导丝不会影响微透析的体外回收率。这些是皮肤体内微透析研究中最常用的流速。