一种用于测定铜蓝蛋白的新型奥林巴斯检测法。

A new Olympus assay for the determination of ceruloplasmin.

作者信息

Larkin Tracey C, Fingleton Teresina A, Mccusker Matthew, Cassidy Martin, Gunzer Gerhard, Lepp Cyrus A, Gamboa Erby

机构信息

Olympus Diagnostica GmbH (Ireland), Lismeehan, O'Callaghans Mills, Co. Clare, Ireland.

出版信息

Clin Lab. 2004;50(3-4):193-203.

Abstract

Ceruloplasmin's main clinical importance is in the diagnosis of Wilson's disease, where plasma ceruloplasmin concentration is reduced. Increased levels are particularly notable in diseases of the reticuloendothelial system such as Hodgkin's disease as well as during pregnancy or use of contraceptive pills. This paper describes a new Olympus assay OSR6164 for the determination of ceruloplasmin on the Olympus AU640E, AU400E and AU2700 analytical systems. The turbidimetric assay has a linear measurable range of 60-2000 mg/l. Prozone hook effect greater than 9000 mg/l. Imprecision CV values < 5% for within-run and < 10% for between-day. Lipemic, hemolytic and icteric interference of < or = 5%. Traceability of recovery to CRM470 standardisation. Method comparison to Behring Nephelometry yielded a correlation (r) of 0.99 and a slope of 1.02. On-board and calibration stability of 60 days and 14 days, respectively. Serumor plasma (EDTA and lithium heparin anticoagulants) samples may be used.

摘要

铜蓝蛋白的主要临床意义在于诊断威尔逊病,该病患者血浆铜蓝蛋白浓度降低。在网状内皮系统疾病(如霍奇金病)以及孕期或服用避孕药期间,铜蓝蛋白水平会升高。本文介绍了一种用于在奥林巴斯AU640E、AU400E和AU2700分析系统上测定铜蓝蛋白的新型奥林巴斯检测方法OSR6164。该比浊法的线性可测量范围为60 - 2000mg/l。前带钩效应大于9000mg/l。批内不精密度CV值<5%,批间<10%。脂血、溶血和黄疸干扰≤5%。回收率可溯源至CRM470标准品。与贝林比浊法的方法比较得出相关性(r)为0.99,斜率为1.02。试剂在分析仪上的稳定性为60天,校准稳定性为14天。可使用血清或血浆(EDTA和肝素锂抗凝剂)样本。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验