Localization of nerve depolarization with magnetic stimulation.

作者信息

Odderson I R, Halar E M

机构信息

Department of Rehabilitation Medicine, University of Washington School of Medicine, Seattle 98195.

出版信息

Muscle Nerve. 1992 Jun;15(6):711-5. doi: 10.1002/mus.880150614.

Abstract

The specific location on the magnetic stimulation (MS) coil that may correspond to the area of nerve depolarization has not been determined. In order to localize such an area, MS with 9-cm and 5-cm diameter coils was compared with conventional percutaneous electric stimulation (ES). On the 9-cm coil the distribution of points of nerve depolarization corresponded to that quarter of the coil which was placed over and parallel to the median nerve, whereas on the 5-cm coil, this area also extended outside the coil. The points of median nerve depolarization with MS were distributed over a distance of 7 cm on the stimulator head and was nearly identical for the 2 coil sizes at the wrist and elbow. Ulnar nerve costimulation was less frequent with the smaller coil at the wrist. A calculated reference point on the coil is suggested for more accurate NCV determinations.

摘要

磁刺激(MS)线圈上可能与神经去极化区域相对应的具体位置尚未确定。为了定位这样一个区域,将直径为9厘米和5厘米的线圈进行磁刺激与传统经皮电刺激(ES)进行了比较。在9厘米的线圈上,神经去极化点的分布与放置在正中神经上方并与之平行的线圈的四分之一相对应,而在5厘米的线圈上,该区域也延伸到了线圈之外。磁刺激引起正中神经去极化的点在刺激器头部的分布距离为7厘米,在手腕和肘部,两种线圈尺寸的情况几乎相同。在手腕处,较小的线圈进行尺神经共刺激的频率较低。建议在线圈上计算一个参考点,以更准确地测定神经传导速度(NCV)。