Kurata A, Irikura K, Miyasaka Y, Yada K, Kan S
Department of Neurosurgery, Kitasato University School of Medicine.
No Shinkei Geka. 1992 Aug;20(8):849-56.
BALT magic catheter (PURSIL catheter) is a new catheter which is more flexible and plaint than the TRACKER-18 catheter because its mid-section and distal portion are made from polyurethane and silicon rather than polyethylene. Because the PURSIL catheter is flexible, it cannot be inserted with a guide wire, although it can be inserted into circulation during injection of contrast medium. Its inner diameter, 3.5mm, is smaller than that (5mm) of the TRACKER. There are few reports of embolization with the PURSIL catheter, and these describe only embolization using adhesive liquid embolus material. Embolization using small particles through the PURSIL catheter has never been reported. We report the catheterization using the PURSIL catheter of 26 arteries in 15 patients. The catheter was used for superselective angiography to clarify feeder and amytal tests in 5 patients with AVM, and embolization using PVA particles (150-250 microns) was performed in 1 patient with AVM, 3 patients with dural AVM, 4 cases with meningioma and 1 patient with vertebral aneurysm. Small PVA particles were easily injected without resistance through this catheter. In only one patient with dural AVM, was catheterization with this catheter not successful, and that was because the distal portion of the internal maxillar artery was extremely tortuous because of severe atherosclerosis. No complications associated with the use of this catheter were observed. We think that the greatest advantage of this catheter is the safety afforded by not using a guide wire. A disadvantage of this catheter is that, because of its small inner diameter, only small particles, or liquid embolus material can be injected through it.(ABSTRACT TRUNCATED AT 250 WORDS)
BALT神奇导管(PURSIL导管)是一种新型导管,比TRACKER - 18导管更柔软且更顺应性,因为其中段和远端部分由聚氨酯和硅制成,而非聚乙烯。由于PURSIL导管很柔软,它不能通过导丝插入,尽管在注射造影剂时可插入循环系统。其内径为3.5毫米,小于TRACKER导管的内径(5毫米)。关于使用PURSIL导管进行栓塞的报道很少,且这些报道仅描述了使用粘性液体栓塞材料的栓塞情况。从未有过通过PURSIL导管使用小颗粒进行栓塞的报道。我们报告了对15例患者的26条动脉使用PURSIL导管进行插管的情况。该导管用于5例动静脉畸形(AVM)患者的超选择性血管造影以明确供血动脉和阿米妥试验,对1例AVM患者、3例硬脑膜AVM患者、4例脑膜瘤患者和1例椎动脉动脉瘤患者使用聚乙烯醇(PVA)颗粒(150 - 250微米)进行栓塞。小PVA颗粒通过该导管易于无阻力地注入。仅1例硬脑膜AVM患者使用该导管插管未成功,原因是上颌内动脉远端因严重动脉粥样硬化而极度迂曲。未观察到与使用该导管相关的并发症。我们认为该导管最大的优点是不使用导丝所带来的安全性。该导管的一个缺点是,由于其内径小,只能通过它注入小颗粒或液体栓塞材料。(摘要截断于250字)