Suppr超能文献

用于MRI引导前列腺热疗的具有旋转控制功能的高定向经尿道超声探头。

Highly directional transurethral ultrasound applicators with rotational control for MRI-guided prostatic thermal therapy.

作者信息

Ross Anthony B, Diederich Chris J, Nau William H, Gill Harcharan, Bouley Donna M, Daniel Bruce, Rieke Viola, Butts R Kim, Sommer Graham

机构信息

Thermal Therapy Research Group, UCSF Radiation Oncology, San Francisco, CA, USA.

出版信息

Phys Med Biol. 2004 Jan 21;49(2):189-204. doi: 10.1088/0031-9155/49/2/002.

Abstract

Transurethral ultrasound applicators with highly directional energy deposition and rotational control were investigated for precise treatment of benign prostatic hyperplasia (BPH) and adenocarcinoma of the prostate (CaP). Two types of catheter-based applicators were fabricated, using either 90 degrees sectored tubular (3.5 mm OD x 10 mm) or planar transducers (3.5 mm x 10 mm). They were constructed to be MRI compatible, minimally invasive and allow for manual rotation of the transducer array within a 10 mm cooling balloon. In vivo evaluations of the applicators were performed in canine prostates (n = 3) using MRI guidance (0.5 T interventional magnet). MR temperature imaging (MRTI) utilizing the proton resonance frequency shift method was used to acquire multiple-slice temperature overlays in real time for monitoring and guiding the thermal treatments. Post-treatment T1-weighted contrast-enhanced imaging and triphenyl tetrazolium chloride stained tissue sections were used to define regions of tissue coagulation. Single sonications with the 90 degrees tubular applicator (9-15 W, 12 min, 8 MHz) produced coagulated zones covering an 80 degrees wedge of the prostate extending from 1-2 mm outside the urethra to the outer boundary of the gland (16 mm radial coagulation). Single sonications with the planar applicator (15-20 W, 10 min, approximately 8 MHz) generated thermal lesions of approximately 30 degrees extending to the prostate boundary. Multiple sequential sonications (sweeping) of a planar applicator (12 W with eight rotations of 30 degrees each) demonstrated controllable coagulation of a 270 degrees contiguous section of the prostate extending to the capsule boundary. The feasibility of using highly directional transurethral ultrasound applicators with rotational capabilities to selectively coagulate regions of the prostate while monitoring and controlling the treatments with MRTI was demonstrated in this study.

摘要

研究了具有高定向能量沉积和旋转控制功能的经尿道超声探头,用于前列腺增生(BPH)和前列腺腺癌(CaP)的精确治疗。制作了两种基于导管的探头,一种使用90度扇形管状(外径3.5mm×10mm)换能器,另一种使用平面换能器(3.5mm×10mm)。这些探头被制作成与MRI兼容、微创,并允许换能器阵列在10mm冷却球囊内手动旋转。在犬前列腺(n = 3)中使用MRI引导(0.5T介入磁体)对探头进行体内评估。利用质子共振频率偏移法的MR温度成像(MRTI)用于实时获取多层温度叠加图,以监测和引导热治疗。治疗后T1加权对比增强成像和氯化三苯基四氮唑染色的组织切片用于确定组织凝固区域。使用90度管状探头进行单次超声处理(9 - 15W,12分钟,8MHz)产生的凝固区覆盖前列腺80度楔形区域,从尿道外1 - 2mm延伸至腺体的外边界(径向凝固16mm)。使用平面探头进行单次超声处理(15 - 20W,10分钟,约8MHz)产生的热损伤延伸至前列腺边界约30度。对平面探头进行多次连续超声处理(扫描)(12W,每次旋转30度,共8次)显示可控制前列腺270度连续区域的凝固,直至包膜边界。本研究证明了使用具有旋转能力的高定向经尿道超声探头在通过MRTI监测和控制治疗的同时选择性凝固前列腺区域的可行性。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验