Eyding Jens, Wilkening Wilko, Reckhardt Markus, Meves Saskia, Postert Thomas
Department of Neurology, St. Josef University Hospital, Ruhr-University Bochum, Gudrunstr. 56, D-44791 Bochum, Germany.
J Neuroimaging. 2004 Apr;14(2):143-9.
Contrast burst depletion imaging (CODIM) visualizes cerebral perfusion by destruction of microbubbles and observation of image intensity course. Because of its complexity, artifacts occur. Criteria of reliability to improve diagnostic significance were created and validated.
Eighteen healthy volunteers were examined with 2 echo contrast agents (ECAs) and 3 frame rates in 3 regions of interest (ROIs). Perfusion coefficient (PC), Tmin (time to decrease intensity to 10% of its max), and relative error (RE) (deviation of measured data from fitted model) were determined. PC differed significantly neither between CA nor between frame rates (overall mean = 1.60 +/- 0.21 x 10(-2) s-1). Tmin differed significantly between frame rate groups (P < .001, 33.4 +/- 11.2 s/0.5 Hz; 3.6 +/- 2.5 s/5 Hz) since it is related to destruction of microbubbles that occurs with each frame and to the perfusion rate. RE was higher in the Optison group and tended to decrease in ROIs closer to the probe.
PC was independent of frame rate and ECA. Tmin was shorter with higher frame rates. Due to a very rapid decay at 5 Hz, the ideal frame rate should be about 1 Hz, that is, because the number of frames acquired within Tmin and therefore signal-to-noise ratio is higher at 1 Hz. Since the algorithm is complex (high RE) and more artifacts should occur in patients (insufficient bone window, etc), a triggering of the insonations by, for example, heart rate could decrease artifacts and increase diagnostic power of CODIM.
对比剂猝灭成像(CODIM)通过微泡破坏及观察图像强度变化过程来显示脑灌注情况。因其操作复杂,会出现伪影。为此制定并验证了提高诊断意义的可靠性标准。
对18名健康志愿者在3个感兴趣区域(ROI)使用2种超声造影剂(ECA)及3种帧率进行检查。测定灌注系数(PC)、Tmin(强度降至最大值10%的时间)和相对误差(RE)(测量数据与拟合模型的偏差)。PC在不同造影剂及不同帧率之间均无显著差异(总体均值 = 1.60 +/- 0.21 x 10(-2) s-1)。Tmin在不同帧率组间差异显著(P <.001,0.5 Hz时为33.4 +/- 11.2 s;5 Hz时为3.6 +/- 2.5 s),因为它与每帧发生的微泡破坏及灌注速率有关。Optison组的RE较高,且在靠近探头的ROI中RE有降低趋势。
PC与帧率和ECA无关。帧率越高,Tmin越短。由于在5 Hz时衰减非常快,理想帧率应为约1 Hz,这是因为在Tmin内采集的帧数以及因此的信噪比在1 Hz时更高。由于该算法复杂(RE高)且在患者中可能出现更多伪影(骨窗不足等),例如通过心率触发超声照射可减少伪影并提高CODIM的诊断能力。