脉搏血氧仪探头的对比测试。

Comparative testing of pulse oximeter probes.

作者信息

van Oostrom Johannes H, Melker Richard J

机构信息

Department of Anesthesiology, College of Medicine, and the McKnight Brain Institute, University of Florida, Gainesville, Florida 32610-0254, USA.

出版信息

Anesth Analg. 2004 May;98(5):1354-8, table of contents. doi: 10.1213/01.ane.0000105869.91626.3a.

Abstract

UNLABELLED

The testing of pulse oximeter probes is generally limited to the integrity of the electrical circuit and does not include the optical properties of the probes. Few pulse oximeter testers evaluate the accuracy of both the monitor and the probe. We designed a study to compare the accuracy of nonproprietary probes (OSS Medical) designed for use with Nellcor, Datex-Ohmeda, and Criticare pulse oximeter monitors with that of their corresponding proprietary probes by using a commercial off-the-shelf pulse oximeter tester (Index). The Index pulse oximeter tester does include testing of the optical properties of the pulse oximeter probes. The pulse oximeter tester was given a controlled input that simulated acute apnea. Desaturation curves were automatically recorded from the pulse oximeter monitors with a data-collection computer. Comparisons between equivalent proprietary and nonproprietary probes were performed. Data were analyzed by using univariate and multivariate general linear model analysis. Five OSS Medical probe models were statistically better than the equivalent proprietary probes. The remainder of the probes were statistically similar. Comparative and simulation studies can have significant advantages over human studies because they are cost-effective, evaluate equipment in a clinically relevant scenario, and pose no risk to patients, but they are limited by the realism of the simulation.

IMPLICATIONS

We studied the performance of pulse oximeter probes in a simulated environment. Our results show significant differences between some probes that affect the accuracy of measurement.

摘要

未贴标签

脉搏血氧饱和度仪探头的测试通常仅限于电路的完整性,不包括探头的光学特性。很少有脉搏血氧饱和度仪测试仪能评估监护仪和探头的准确性。我们设计了一项研究,通过使用市售的脉搏血氧饱和度仪测试仪(Index),比较用于Nellcor、Datex - Ohmeda和Criticare脉搏血氧饱和度仪监护仪的非专利探头(OSS Medical)与其相应专利探头的准确性。Index脉搏血氧饱和度仪测试仪确实包括对脉搏血氧饱和度仪探头光学特性的测试。给脉搏血氧饱和度仪测试仪一个模拟急性呼吸暂停的受控输入。通过数据采集计算机自动记录脉搏血氧饱和度仪监护仪的去饱和曲线。对等效的专利探头和非专利探头进行了比较。数据采用单变量和多变量一般线性模型分析。五种OSS Medical探头型号在统计学上优于等效的专利探头。其余探头在统计学上相似。比较和模拟研究相对于人体研究可能具有显著优势,因为它们具有成本效益,能在临床相关场景中评估设备,且对患者无风险,但它们受模拟逼真度的限制。

启示

我们在模拟环境中研究了脉搏血氧饱和度仪探头的性能。我们的结果显示,一些探头之间存在显著差异,这会影响测量的准确性。