Jones J G
University Department of Anaesthesia, Addenbrooke's Hospital, Hills Road, Cambridge, CB2 2QQ, UK.
Anaesthesia. 2004 Jul;59(7):704-9. doi: 10.1111/j.1365-2044.2004.03754.x.
Fixed performance venturi devices should provide a predetermined oxygen concentration at an outflow which exceeds an adult's peak resting inspiratory flow rate (approximately 30 l.min(-1)). Campbell's original description mentioned the sensitivity of the venturi device to downstream resistance but gave no further details. This study examined outflow and oxygen concentration from the five standard venturi devices (24-60% O(2)) when downstream pressure increased. Outflow was exquisitely sensitive to small increases in pressure. The outflow at zero downstream pressure for the 24-40% O(2) venturi devices ranged from 40 to 50 l.min(-1) but only 2-3 mmH(2)O was needed to halve this flow and increase oxygen concentration. The 60% O(2) venturi delivered a maximum of only 30 l.min(-1) at zero downstream pressure and flow was reduced further by increasing this pressure. An increase in downstream pressure of only a few mmH(2)O increased oxygen concentration and decreased outflow of all the venturi devices tested, in most to less than normal peak tidal flow in adults.
固定性能的文丘里装置应在流出端提供预定的氧浓度,该浓度要超过成年人静息时的最大吸气流量(约30升·分钟⁻¹)。坎贝尔最初的描述提到了文丘里装置对下游阻力的敏感性,但未给出更多细节。本研究考察了当下游压力增加时,五种标准文丘里装置(氧浓度为24% - 60%)的流出量和氧浓度。流出量对压力的小幅增加极为敏感。24% - 40%氧浓度的文丘里装置在下游压力为零时的流出量范围为40至50升·分钟⁻¹,但只需2 - 3毫米水柱的压力就能使该流量减半并提高氧浓度。60%氧浓度的文丘里装置在下游压力为零时的最大流出量仅为30升·分钟⁻¹,且随着压力增加流量会进一步降低。仅几毫米水柱的下游压力增加就会提高所有测试文丘里装置的氧浓度并降低其流出量,在大多数情况下,流出量会降至低于成年人正常的潮气量峰值。