Musshoff F, Daldrup T
Institut für Rechtsmedizin, Heinrich-Heine-Universität Düsseldorf, Federal Republic of Germany.
Int J Legal Med. 1992;105(2):105-9. doi: 10.1007/BF02340833.
A rapid and quantitative method for the determination of benzodiazepines using high-performance liquid chromatography (HPLC) with diode-array detection (DAD) is reported. The drugs were extracted from serum, blood or post-mortem blood using C18 extraction columns. Brotizolam was used as internal standard. Experiments with spiked serum/blood samples resulted in recoveries between 75% and 94% for all investigated benzodiazepines. Excellent linearity was obtained over the concentration range 5-1500 ng benzodiazepine/ml. The limit of detection was approximately 2 ng/ml. The detection of low therapeutic serum levels of highly potent benzodiazepines is also possible.