Barros Marcílio G, Osher Robert H
Cincinnati Eye Institute, 10494 Montgomery Road, Cincinnati, OH 45242, USA.
J Cataract Refract Surg. 2004 Jul;30(7):1570-1. doi: 10.1016/j.jcrs.2003.11.059.
We describe a surgical complication with the Alcon silicone irrigation/aspiration (I/A) tip. In 2 consecutive cases, the posterior capsule was torn because of a misalignment between the inner metallic shaft and the hole in the silicone I/A tip. To achieve the protective benefits of this device, the surgeon must visually confirm proper alignment of the outer silicone and inner metallic portion of the tip.
我们描述了一种与爱尔康硅胶冲洗/抽吸(I/A)头相关的手术并发症。在连续2例病例中,由于内部金属杆与硅胶I/A头的孔未对齐,导致后囊膜撕裂。为了实现该装置的保护作用,外科医生必须目视确认该I/A头外部硅胶部分与内部金属部分的正确对齐。