通过使用五种不同的牙合架进行测试,对电子面弓的可靠性和有效性进行体外评估。

An in vitro evaluation of the reliability and validity of an electronic pantograph by testing with five different articulators.

作者信息

Chang Willy S W, Romberg Elaine, Driscoll Carl F, Tabacco Michael J

机构信息

Department of Restorative Dentistry, Baltimore College of Dental Surgery, Dental School, University of Maryland, 21201, USA.

出版信息

J Prosthet Dent. 2004 Jul;92(1):83-9. doi: 10.1016/j.prosdent.2004.04.011.

Abstract

STATEMENT OF PROBLEM

The degree of reliability and validity of a new electronic pantographic instrument, the Cadiax Compact, has not been established.

PURPOSE

The purpose of this study was to test the reliability and validity of the electronic pantograph in calculating condylar settings for 5 different articulators (Denar D5A, Denar Mark II, Whip Mix 8500, Hanau Modular, and Panadent PCH).

MATERIAL AND METHODS

Pantograph sensors were mounted to each articulator with custom-made mounting devices. Border movements were made on each articulator with known condylar settings to produce readings at 3-, 5-, and 10-mm condylotrack distances. The condylar settings investigated included horizontal condylar inclination (HCI), immediate mandibular lateral translation (IMLT), progressive mandibular lateral translation (PMLT), top wall, and rear wall. Reliability was assessed by the relative size of standard deviations. Validity was tested with 1-way analysis of variance and the Tukey HSD Test (alpha=.05).

RESULTS

The reliability readings for the condylar settings at the 10-mm condylotrack distance, for the most part, were more consistent than those at the 5-mm distance, which, in turn, were more consistent than those at the 3-mm distance. When there were exceptions to this trend, the differences in standard deviations were very small (0.006 mm for IMLT, 0.03 and 0.12 degrees for HCI). When comparing validity between the condylotrack distances, the smallest deviations, in general, were found at the 10-mm distance, the next smallest at the 5-mm distance, and the largest, at the 3-mm distance for HCI. For PMLT, in general, the 10-mm and 5-mm deviations were smaller than the 3-mm deviation. For IMLT, there was no significant difference between the 3-, 5-, and 10-mm deviations. In analyzing validity between the articulators, the smallest deviations from the preset values, in general, were found with the Denar Mark II.

CONCLUSION

The standard deviations for assessing reliability and the mean deviations for assessing validity were both relatively small in comparison to the average values of the condylar determinants. Therefore, the electronic pantograph was determined to be both reliable and valid.

摘要

问题陈述

一种新型电子全景描记仪Cadiax Compact的可靠性和有效性程度尚未确定。

目的

本研究的目的是测试该电子全景描记仪在为5种不同牙合架(Denar D5A、Denar Mark II、Whip Mix 8500、Hanau Modular和Panadent PCH)计算髁突位置时的可靠性和有效性。

材料与方法

使用定制的安装装置将全景描记仪传感器安装到每个牙合架上。在每个牙合架上进行边缘运动,采用已知的髁突位置,以获取在髁突轨迹距离为3毫米、5毫米和10毫米时的读数。所研究的髁突位置包括水平髁突倾斜度(HCI)、即刻下颌侧方移位(IMLT)、渐进性下颌侧方移位(PMLT)、顶壁和后壁。通过标准差的相对大小评估可靠性。采用单因素方差分析和Tukey HSD检验(α = 0.05)测试有效性。

结果

在髁突轨迹距离为10毫米时,髁突位置的可靠性读数在大多数情况下比5毫米距离时更一致,而5毫米距离时又比3毫米距离时更一致。当出现与该趋势不符的情况时,标准差的差异非常小(IMLT为0.006毫米,HCI为0.03度和0.12度)。在比较不同髁突轨迹距离之间的有效性时,一般在10毫米距离处偏差最小,5毫米距离处次之,3毫米距离处的HCI偏差最大。对于PMLT,一般10毫米和5毫米的偏差小于3毫米的偏差。对于IMLT,3毫米、5毫米和l0毫米的偏差之间无显著差异。在分析不同牙合架之间的有效性时,一般与预设值偏差最小的是Denar Mark II。

结论

与髁突决定因素的平均值相比,用于评估可靠性的标准差和用于评估有效性的平均偏差都相对较小。因此,该电子全景描记仪被确定为既可靠又有效。