Suppr超能文献

用于毛细管电泳-电喷雾电离-质谱分析的斜面尖端鞘液接口。

A beveled tip sheath liquid interface for capillary electrophoresis-electrospray ionization-mass spectrometry.

作者信息

Tseng Mei-Chun, Chen Yet-Ran, Her Guor-Rong

机构信息

Department of Chemistry, National Taiwan University, Taipei, Taiwan, R.O.C.

出版信息

Electrophoresis. 2004 Jul;25(13):2084-2089. doi: 10.1002/elps.200305907.

Abstract

A simple and durable sheath liquid interface for capillary zone electrophoresis-electrospray ionization-mass spectrometry (CZE-ESI-MS) has been developed. This interface utilized a beveled tip emitter and was found to be more sensitive than the conventional sheath liquid interface. The use of a beveled tip reduces the optimal flow rate and therefore decreases sample dilution. The interface utilized a 380 microm inner diameter and 400 microm outer diameter beveled tapered tip. Because of the large inner diameter and outer diameter of the tip, the interface is robust and can be easily implemented. The performance of this interface for CZE-ESI-MS and micelle electrokinetic capillary electrophoresis-electrospray-mass spectrometry, as demonstrated by the analysis of synthetic drugs and triazine mixtures, was significantly better than results obtained using a conventional sheath liquid interface.

摘要

已开发出一种用于毛细管区带电泳-电喷雾电离-质谱联用(CZE-ESI-MS)的简单耐用的鞘液接口。该接口采用了斜面尖端发射器,发现其比传统鞘液接口更灵敏。使用斜面尖端可降低最佳流速,从而减少样品稀释。该接口采用了内径380微米、外径400微米的斜面锥形尖端。由于尖端的内径和外径较大,该接口坚固耐用且易于实现。通过对合成药物和三嗪混合物的分析表明,该接口用于CZE-ESI-MS和胶束电动毛细管电泳-电喷雾质谱联用的性能明显优于使用传统鞘液接口获得的结果。