Suppr超能文献

Source configuration and dose rates for the Selectron afterloading equipment for gynecologic applicators.

作者信息

Grigsby P W, Williamson J F, Perez C A

机构信息

Mallinckrodt Institute of Radiology, Washington University School of Medicine, St. Louis, MO.

出版信息

Int J Radiat Oncol Biol Phys. 1992;24(2):321-7. doi: 10.1016/0360-3016(92)90688-e.

Abstract

This paper describes source train configurations of the Selectron LDR-6 that can be used to duplicate commonly used manually afterloadable source configurations for gynecological intracavitary brachytherapy. We demonstrate that the recommended source configuration using 2.5 mm 137Cs spheres is dosimetrically equivalent to its manually afterloaded counterparts using 3M 137Cs sources (20 mm overall length, 14 mm active length). Isodose distributions are in agreement to within 5% for the Fletcher-Suit-Delclos family of tandems, colpostats, and vaginal cylinders. Our data also demonstrate that the isotropic point source model used by most computerized treatment planning systems is a good approximation to source trains of filtered, spherical sources interspersed with solid steel spacers. We conclude that the use of 2.5 mm 137Cs sources in the Selectron remote afterloading system can accurately reproduce the isodose distribution achieved with linear 137Cs manually afterloaded sources.