Peres Osmair, Oliveira Celso H, Barrientos-Astigarraga Rafael E, Rezende Vinícius M, Mendes Gustavo D, de Nucci Gilberto
Department of Pharmacology, State University of Campinas, Campinas, SP, Brazil.
Arzneimittelforschung. 2004;54(6):314-9. doi: 10.1055/s-0031-1296977.
An analytical method based on liquid chromatography with positive ion electrospray ionization (ESI) coupled to tandem mass spectrometry detection was developed for the determination of pantoprazole (CAS 102625-70-7) in human plasma using lansoprazole (CAS 103577-45-3) as the internal standard. The analyte and internal standard were extracted from the plasma samples by liquid/liquid extraction using diethyl-ether/dichloromethane (70:30; v/v) and chromatographed on a C8 analytical column. The mobile phase consisted of acetonitrile/ water/methanol (57:25:18; v/v/v) + 10 mmol/l acetic acid + 20 mmol/l ammonium acetate. The method has a chromatographic total run time of 4.5 min and was linear within the range 5.0-5,000 ng/ mL. Detection was performed on a triple quadrupole tandem mass spectrometer by Multiple Reaction Monitoring (MRM). The intra- and inter-run precisions calculated from quality control (QC) samples were 4.2 % and 3.2 %, respectively. The accuracies as determined from QC samples were -5.0 % (intra-run) and 2.0 % (inter-run). The method herein described was employed in a bioequivalence study of two tablet formulations of pantoprazole.
建立了一种基于液相色谱-正离子电喷雾电离(ESI)与串联质谱检测联用的分析方法,以兰索拉唑(CAS 103577-45-3)为内标物,用于测定人血浆中的泮托拉唑(CAS 102625-70-7)。采用乙醚/二氯甲烷(70:30;v/v)通过液-液萃取从血浆样品中提取分析物和内标物,并在C8分析柱上进行色谱分离。流动相由乙腈/水/甲醇(57:25:18;v/v/v)+10 mmol/L乙酸+20 mmol/L乙酸铵组成。该方法的色谱总运行时间为4.5分钟,在5.0-5000 ng/mL范围内呈线性。在三重四极杆串联质谱仪上通过多反应监测(MRM)进行检测。由质量控制(QC)样品计算得出的批内和批间精密度分别为4.2%和3.2%。由QC样品确定的准确度为批内-5.0%和批间2.0%。本文所述方法用于泮托拉唑两种片剂剂型的生物等效性研究。