Xie Zhiyong, Chen Xiaoyan, Jin Fengdan, Zhong Dafang
Laboratory of Drug Metabolism and Pharmacokinetics, Shenyang Pharmaceutical University, 103 Wenhua Road, Shenyang 110016, P.R. China.
J Chromatogr Sci. 2005 May-Jun;43(5):271-5. doi: 10.1093/chromsci/43.5.271.
A simple and sensitive high-performance liquid chromatography (HPLC) method is developed and validated for simultaneous determination of pantoprazole and its two metabolites (pantoprazole sulfone and pantoprazole thioether) in dog plasma and applied to a pharmacokinetic study in Beagle dogs. Following a protein precipitation procedure, the samples are separated using reversed-phase HPLC (C18) by a gradient of acetonitrile and ammonium acetate (pH 6.0) at a flow rate of 1.0 mL/min and quantitated using UV detection at 290 nm. Omeprazole is selected as the internal standard. The method has a lower limit of quantitation of 0.025 microg/mL for pantoprazole and its two metabolites, using 0.1-mL aliquots of plasma. The linear calibration curves are obtained in the concentration range of 0.025-10.0 microg/mL for three analytes. The intra- and interrun precision (relative standard deviation), calculated from quality control (QC) samples, is less than 13% for three analytes. The accuracy determined from QC samples is between -6.4% and 12%.
建立了一种简单、灵敏的高效液相色谱(HPLC)方法,用于同时测定犬血浆中泮托拉唑及其两种代谢产物(泮托拉唑砜和泮托拉唑硫醚),并将其应用于比格犬的药代动力学研究。经过蛋白沉淀程序后,样品采用反相HPLC(C18),以乙腈和醋酸铵(pH 6.0)梯度洗脱,流速为1.0 mL/min进行分离,并在290 nm处采用紫外检测进行定量。选用奥美拉唑作为内标。该方法使用0.1 mL血浆等分试样时,泮托拉唑及其两种代谢产物的定量下限为0.025 μg/mL。三种分析物在0.025 - 10.0 μg/mL浓度范围内获得线性校准曲线。根据质量控制(QC)样品计算,三种分析物的批内和批间精密度(相对标准偏差)均小于13%。由QC样品确定的准确度在-6.4%至12%之间。