Fetal ventricular mass determination on three-dimensional echocardiography: studies in normal fetuses and validation experiments.

作者信息

Bhat Aarti Hejmadi, Corbett Virginia, Carpenter Nathan, Liu Nick, Liu Ruolan, Wu Anna, Hopkins Graham, Sohaey Roya, Winkler Carey, Sahn Christine S, Sovinsky Valerie, Li Xiaokui, Sahn David J

机构信息

Clinical Care Center for Congenital Heart Disease, Oregon Health & Science University, Portland 97239-3098, USA.

出版信息

Circulation. 2004 Aug 31;110(9):1054-60. doi: 10.1161/01.CIR.0000139848.33468.22. Epub 2004 Aug 23.

Abstract

BACKGROUND

Estimation of ventricular volume and mass is important for baseline and serial evaluation of fetuses with normal or abnormal hearts. Direct measurement of chamber wall volumes and mass can be made without geometric assumptions by 3D fetal echocardiography. Our goals were to determine the feasibility of using fast nongated 3D echocardiography for fetal volumetric and mass assessments, to validate the accuracy of the ultrasound system and the measurement technique, and if satisfactory, to develop normal values for fetal ventricular mass during the second and third trimesters.

METHODS AND RESULTS

This was a prospective outpatient study of 90 consecutive normal pregnancies during routine obstetric services at Oregon Health & Science University (Portland). Optimized 3D volumes of the fetal thorax and cardiac chambers were rapidly acquired and later analyzed for right and left ventricular mass by radial summation technique from manual epicardial and endocardial traces. Experiments to validate the ultrasound system and measurement technique were performed with modified small balloon models and in vivo and ex vivo small animal experiments. Our study established the feasibility of fetal ventricular mass measurements with 3D ultrasound technology and developed normal values for right and left ventricular mass from 15 weeks' gestation to term.

CONCLUSIONS

Nongated fast 3D fetal echocardiography is an acceptable modality for determination of cardiac chamber wall volume and mass with good accuracy and acceptable interobserver variability. The method should be especially valuable as an objective serial measurement in clinical fetal studies with structurally or functionally abnormal hearts.

摘要

背景

心室容积和质量的评估对于正常或异常心脏胎儿的基线和系列评估至关重要。三维胎儿超声心动图可在不做几何假设的情况下直接测量心腔壁容积和质量。我们的目标是确定使用快速非门控三维超声心动图进行胎儿容积和质量评估的可行性,验证超声系统和测量技术的准确性,若结果令人满意,则制定孕中期和孕晚期胎儿心室质量的正常值。

方法与结果

这是一项前瞻性门诊研究,对俄勒冈健康与科学大学(波特兰)常规产科服务期间连续90例正常妊娠进行了研究。快速采集胎儿胸部和心腔的优化三维容积,随后通过手动心外膜和心内膜轨迹的径向求和技术分析右心室和左心室质量。使用改良的小气球模型以及体内和体外小动物实验进行了验证超声系统和测量技术的实验。我们的研究确立了使用三维超声技术测量胎儿心室质量可行性,并制定了从妊娠15周直至足月的右心室和左心室质量正常值。

结论

非门控快速三维胎儿超声心动图是一种可接受的方法,用于确定心腔壁容积和质量,具有良好的准确性和可接受的观察者间变异性。该方法作为对结构或功能异常心脏的临床胎儿研究中的客观系列测量应特别有价值。