Smith D L, Walinsky P, Martinez-Hernandez A, Rosen A, Sterzer F, Kosman Z
Thomas Jefferson University Hospital, Philadelphia, PA 19107.
Am Heart J. 1992 Jun;123(6):1516-21. doi: 10.1016/0002-8703(92)90803-4.
The feasibility of utilizing microwave energy for thermal angioplasty was evaluated in 28 normal New Zealand white rabbits. Angioplasty catheters were inserted via bilateral femoral arteriotomy and balloon angioplasty was performed on the external iliac arteries. There were three variables in the microwave angioplasty procedure: peak temperature, energy duration, and peak balloon pressure. At 1 week after the angioplasty, animals were put to death and the iliac arteries were removed after perfusion fixation. Histologic analysis was performed. Injuries to the medial and intimal proliferation were evaluated using an index based on the circumferential extent and depth of abnormality. There was no relationship noted between intimal proliferation or medial injury and energy duration of peak balloon pressure. There was a direct relationship between peak temperature and medial injury. Furthermore, there was an inverse relationship between medial injury and intimal proliferation. The axial extent of change induced by microwave energy was 1.6 cm. Microwave thermal energy delivery through a balloon angioplasty catheter is feasible and can be performed conveniently.
在28只正常的新西兰白兔身上评估了利用微波能量进行热血管成形术的可行性。通过双侧股动脉切开术插入血管成形术导管,并对双侧髂外动脉进行球囊血管成形术。微波血管成形术过程中有三个变量:峰值温度、能量持续时间和峰值球囊压力。血管成形术后1周,处死动物,在灌注固定后取出髂动脉。进行组织学分析。使用基于异常的周向范围和深度的指数评估中膜损伤和内膜增生。未发现内膜增生或中膜损伤与峰值球囊压力的能量持续时间之间存在关联。峰值温度与中膜损伤之间存在直接关系。此外,中膜损伤与内膜增生之间存在负相关关系。微波能量引起的变化的轴向范围为1.6厘米。通过球囊血管成形术导管输送微波热能是可行的,并且可以方便地进行。