Suppr超能文献

[椎弓根螺钉置入术中的三维导航]

[Intraoperative three-dimensional navigation for pedicle screw placement].

作者信息

Grützner P A, Beutler T, Wendl K, von Recum J, Wentzensen A, Nolte L-P

机构信息

Berufsgenossenschaftliche Unfallklinik Ludwigshafen, Klinik für Unfall- und Wiederherstellungschirurgie an der Universität Heidelberg.

出版信息

Chirurg. 2004 Oct;75(10):967-75. doi: 10.1007/s00104-004-0944-3.

Abstract

The mobile SIREMOBIL Iso-C(3D) C-arm (Siemens, Erlangen, Germany) is the first device permitting intraoperative, three-dimensional representation of bone structures. A high-resolution, isotropic 3D data cube in the isocenter with sides of approximately 12 cm is calculated simultaneously. The SIREMOBIL Iso-C(3D) is linked to the navigation system. This makes it possible to transfer the generated 3D data directly to the linked navigation system without the need for surgeon-dependent registration. In this prospective clinical trial, we evaluated the accuracy of pedicle screw placement using this device. In 61 patients, a total of 302 pedicle screws were placed. Only in five cases (1.7%) were misplacements of > or =2 mm shown in postoperative control CT. The average fluoroscopy time was 1.28+/-0.56 min, and the average operative duration was 103.26+/-23.3 min. There were no postoperative neurological complications in any of the 30 patients. From these data, we conclude that Iso-C(3D) navigation is a very accurate method for the placement of pedicle screws.

摘要

移动型SIREMOBIL Iso-C(3D) C型臂(德国西门子公司,埃尔朗根)是首款能够在术中对骨骼结构进行三维呈现的设备。同时,在等中心处计算出一个边长约为12厘米的高分辨率、各向同性的三维数据立方体。SIREMOBIL Iso-C(3D)与导航系统相连。这使得生成的三维数据能够直接传输至相连的导航系统,而无需依赖术者进行配准。在这项前瞻性临床试验中,我们评估了使用该设备进行椎弓根螺钉置入的准确性。61例患者共置入302枚椎弓根螺钉。术后对照CT仅显示5例(1.7%)螺钉误置≥2毫米。平均透视时间为1.28±0.56分钟,平均手术时长为103.26±23.3分钟。30例患者中无一例出现术后神经并发症。基于这些数据,我们得出结论,Iso-C(3D)导航是一种非常精确的椎弓根螺钉置入方法。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验