Deflated balloon profile is an important consideration in the selection of PTCA catheters. In order to test the accuracy of manufacturer-reported balloon profiles, we measured deflated shoulder and mid-balloon profile in 107 unused catheters (2.0, 2.5, and 3.0 mm) from 6 manufacturers using a precision hole gauge with a resolution of .001 inches. A significant difference was defined as a discrepancy greater than or equal to 0.003 inches from the manufacturer's reported profile. During Phase I, measurements were obtained at room temperature on catheters directly out of the package. Of the 36 models tested, 21/36 (56%) had measured shoulder profiles that were larger than their reported profile, and 29/36 (81%) had mid-balloon profiles that were larger than their reported profile. During Phase II, measurements were made at body temperature following a 60 second inflation at 120 psi (8.16 atm). Of 33 models tested, 18/33 (55%) had measured shoulder profiles that were larger than their reported profile; 26/33 (79%) had mid-balloon profiles that were larger than their reported profile. The term "profile" is imprecisely defined. Manufacturer-reported deflated PTCA balloon profiles are not always accurate. Independent and uniform testing of balloon profiles is necessary for valid comparisons between balloons and manufacturers.
瘪球囊外形是选择经皮冠状动脉腔内血管成形术(PTCA)导管时的一个重要考量因素。为了测试制造商报告的球囊外形的准确性,我们使用分辨率为0.001英寸的精密孔径规,对来自6家制造商的107根未使用过的导管(2.0、2.5和3.0毫米)的瘪球囊肩部和球囊中部外形进行了测量。显著差异定义为与制造商报告的外形相差大于或等于0.003英寸。在第一阶段,在室温下对刚从包装中取出的导管进行测量。在测试的36个型号中,21/36(56%)测得的肩部外形大于其报告的外形,29/36(81%)测得的球囊中部外形大于其报告的外形。在第二阶段,在120磅力/平方英寸(8.16个大气压)下充气60秒后,在体温下进行测量。在测试的33个型号中,18/33(55%)测得的肩部外形大于其报告的外形;26/33(79%)测得的球囊中部外形大于其报告的外形。术语“外形”定义不精确。制造商报告的瘪PTCA球囊外形并不总是准确的。为了在球囊和制造商之间进行有效的比较,对球囊外形进行独立和统一的测试是必要的。