The relationship of proximal locking screws to the axillary nerve during antegrade humeral nail insertion of four commercially available implants.

作者信息

Prince Edward J, Breien Kristoffer M, Fehringer Edward V, Mormino Matthew A

机构信息

Department of Orthopaedic Surgery & Rehabilitation, University of Nebraska Medical Center, Omaha, NE 68198-1080, USA.

出版信息

J Orthop Trauma. 2004 Oct;18(9):585-8. doi: 10.1097/00005131-200410000-00001.

Abstract

OBJECTIVE

The purpose of this study was to determine the proximity of proximal interlocking mechanisms in 4 current antegrade humeral nails to the axillary nerve and its branches.

DESIGN

Cadaveric study.

SETTING

Anatomy laboratory.

MAIN OUTCOME MEASURE

Anatomic relationships.

METHODS

Four humeral nail designs (labeled SS, SL, SZ, and SN) were each inserted in successive antegrade fashion in 10 cadaveric upper extremity specimens. Three variables were measured: from acromion to the axillary nerve, from acromion to entry sites of proximal locking devices, and from locking devices to axillary nerves and their branches.

RESULTS

In nail SS, the proximally directed oblique locking screw came into contact with the ascending branch of the axillary nerve in 6 of 10 specimens. Mean distance from spiral blades in nails SS and SL were 26 mm to the axillary nerve and 16 mm to its ascending branch. Interlocking screws for nails SZ, SN, and SL did not violate the axillary nerve or its branches in any specimen. Mean distance from lateral acromion to the axillary nerve measured 58.7 mm.

CONCLUSION

Nail SS's oblique locking screw may injure the ascending branch of the axillary nerve. Three of the 4 nails tested did not endanger the axillary nerve. However, when transverse proximal locking screws are inserted from a lateral-to-medial direction, they may endanger an arborized axillary nerve. Blunt dissection should be performed with a visible path to bone before instrumentation to reduce the risk of axillary nerve injury.

摘要

目的

本研究旨在确定4种当前顺行肱骨钉近端锁定机制与腋神经及其分支的接近程度。

设计

尸体研究。

地点

解剖实验室。

主要观察指标

解剖关系。

方法

将4种肱骨钉设计(分别标记为SS、SL、SZ和SN)依次以顺行方式插入10个尸体上肢标本中。测量3个变量:从肩峰到腋神经的距离、从肩峰到近端锁定装置入口点的距离,以及从锁定装置到腋神经及其分支的距离。

结果

在SS型钉中,10个标本中有6个的近端斜向锁定螺钉与腋神经的升支接触。SS型和SL型钉的螺旋刀片到腋神经的平均距离为26 mm,到其升支的平均距离为16 mm。SZ型、SN型和SL型钉的锁定螺钉在任何标本中均未侵犯腋神经及其分支。从肩峰外侧到腋神经的平均距离为58.7 mm。

结论

SS型钉的斜向锁定螺钉可能会损伤腋神经的升支。所测试的4种钉中有3种不会危及腋神经。然而,当从外侧向内侧方向插入横向近端锁定螺钉时,可能会危及分支状的腋神经。在器械操作前,应进行钝性分离并确保有可见的通向骨骼的路径,以降低腋神经损伤的风险。