多平面远端与单平面近端交锁螺钉与神经血管结构的关系:Depuy Synthes 肱骨钉 EX:解剖学研究。

The Relationship of Multiplanar Distal and Uniplanar Proximal Interlocking Screws to Neurovascular Structures with the Depuy Synthes Humeral Nail-EX: An Anatomic Study.

机构信息

Department of Orthopedic Surgery, University of Oklahoma, Oklahoma City, OK.

出版信息

J Orthop Trauma. 2018 Aug;32(8):e300-e303. doi: 10.1097/BOT.0000000000001182.

Abstract

OBJECTIVE

To determine the proximity of neurovascular structures to multiplanar distal interlocking screws and proximal interlocking devices for an antegrade inserted humeral nail.

METHODS

Humeral nails (Depuy Synthes Humeral Nail-EX, Paoli, PA) were inserted into 10 matched cadavers with 3 distal interlocking trajectories: anterior to posterior (A-P), anterolateral to posteromedial, and anteromedial to posterolateral (AM-PL). The distance of the screw path and screw head to the radial, lateral antebrachial cutaneous nerve (LACN), and median nerves and the brachial artery was measured. Proximally, the distance of spiral blades and/or interlocking screws to the axillary nerve was measured.

RESULTS

The median distance of the screw path/screw head inserted (A-P) to the LACN was 0 mm/0 mm in the left arm and 3.5 mm/1.6 mm in the right arm. The median distance of the screw path/screw head inserted AM-PL to the median nerve was 7.08 mm/0 mm in the left arm and 0 mm/0.12 mm in the right arm. The median distance of the screw path/screw head inserted AM-PL to the brachial artery was 4.05 mm/3.83 mm in the left arm and 2.19 mm/2.84 mm in the right arm. The humeral head was radiographically divided into 3 zones. The average distances of the axillary nerve to interlocking devices placed into zone 2 and zone 3 of the humeral head are 14.65 mm and 1.07 mm, respectively.

CONCLUSIONS

When using the DePuy Synthes Humeral Nail-EX, AM-PL screws may risk the median nerve and brachial artery; thus, this trajectory is not recommended. The LACN may be at risk with A-P and to a lesser degree, anterolateral to posteromedial screws. Proximally, interlocking devices in zone 3 of the humeral head may risk the axillary nerve.

摘要

目的

确定多平面远端交锁螺钉和前向插入肱骨钉近端交锁装置与神经血管结构的接近程度。

方法

将肱骨钉(Depuy Synthes 肱骨钉-EX,宾夕法尼亚州波利斯)插入 10 具匹配的尸体中,有 3 种远端交锁轨迹:从前向后(A-P)、前外侧到后内侧,以及前内侧到后外侧(AM-PL)。测量螺钉路径和螺钉头到桡神经、前臂外侧皮神经(LACN)和正中神经以及肱动脉的距离。在近端,测量螺旋刀片和/或交锁螺钉到腋神经的距离。

结果

左侧手臂中,螺钉路径/螺钉头插入(A-P)与 LACN 的中位距离为 0mm/0mm,右侧手臂中为 3.5mm/1.6mm。左侧手臂中,螺钉路径/螺钉头插入 AM-PL 与正中神经的中位距离为 7.08mm/0mm,右侧手臂中为 0mm/0.12mm。左侧手臂中,螺钉路径/螺钉头插入 AM-PL 与肱动脉的中位距离为 4.05mm/3.83mm,右侧手臂中为 2.19mm/2.84mm。肱骨头部被放射状地分为 3 个区。腋神经到放置在肱骨头部 2 区和 3 区的锁定装置的平均距离分别为 14.65mm 和 1.07mm。

结论

使用 DePuy Synthes 肱骨钉-EX 时,AM-PL 螺钉可能会损伤正中神经和肱动脉;因此,不推荐使用这种轨迹。A-P 和前外侧到后内侧螺钉可能会损伤 LACN。在近端,肱骨头部 3 区的锁定装置可能会损伤腋神经。