Abstract

BACKGROUND/AIMS: A new wireless esophageal pH monitoring device, the Bravo system, had been tested in the US with favorable results over conventional systems. However, its clinical application in eastern countries remains limited. The aim of the study was to assess the efficacy of the Bravo system in an Asian population, and compare it with the western results.

METHODOLOGY

Thirty consecutive patients with clinically evident gastroesophageal reflux disease underwent esophageal pH monitoring by using the Bravo system. The Bravo capsule capable of measuring and transmitting pH data was introduced into the squamocolumnar junction and fixed on the esophageal mucosa through a locking pin. All patients were instructed to carry the receiver for a 48-hour study.

RESULTS

Twenty-nine (97%) of the 30 patients were successfully tested. In most cases (26/30, 87%) the capsules were placed without conscious sedation. All patients had successful capsule placement; however, two of them had failed the initial capsule attachment and required a second capsule. All patients had successful 24-hour recordings (30/30, 100%), while the successful 48-hour recording rate was 97% (29/30). A minor mucosal injury caused by inadvertent capsule extraction was the only complication (1/30, 3.3%).

CONCLUSIONS

Compared to the western experience, the new wireless pH monitoring system achieved a comparable attaching rate and recording efficacy. Potential complication was rare and self-limited.

摘要

背景/目的:一种新型无线食管pH监测设备——Bravo系统,已在美国进行测试,与传统系统相比结果良好。然而,其在东方国家的临床应用仍然有限。本研究的目的是评估Bravo系统在亚洲人群中的疗效,并与西方的结果进行比较。

方法

连续30例有明显临床症状的胃食管反流病患者使用Bravo系统进行食管pH监测。将能够测量和传输pH数据的Bravo胶囊置入鳞柱状交界处,并通过锁定针固定在食管黏膜上。所有患者被要求携带接收器进行48小时的研究。

结果

30例患者中有29例(97%)测试成功。在大多数情况下(26/30,87%),胶囊置入时无需清醒镇静。所有患者胶囊置入均成功;然而,其中2例最初胶囊附着失败,需要置入第二个胶囊。所有患者24小时记录均成功(30/30,100%),而48小时成功记录率为97%(29/30)。唯一的并发症是因意外取出胶囊导致的轻微黏膜损伤(1/30,3.3%)。

结论

与西方的经验相比,新型无线pH监测系统的附着率和记录效果相当。潜在并发症罕见且为自限性。