Suppr超能文献

人血清中总甲状腺素测量的计量溯源性:建立参考测量实验室网络的努力。

Metrologic traceability of total thyroxine measurements in human serum: efforts to establish a network of reference measurement laboratories.

作者信息

Thienpont Linda M, Van Uytfanghe Katleen, Marriot John, Stokes Peter, Siekmann Lothar, Kessler Anja, Bunk David, Tai Susan

机构信息

Laboratory for Analytical Chemistry, Faculty of Pharmaceutical Sciences, Ghent University, 9000 Gent, Belgium.

出版信息

Clin Chem. 2005 Jan;51(1):161-8. doi: 10.1373/clinchem.2004.041772. Epub 2004 Nov 18.

Abstract

BACKGROUND

Assuring/demonstrating metrologic traceability of in vitro diagnostics necessitates the availability of measurand-specific reference measurement systems (RMSs) and the possibility for industry to work with competent reference measurement laboratories (RMLs). Here we report the results of a European project to investigate the feasibility of developing a RMS for serum total thyroxine.

METHODS

Four candidate RMLs (cRMLs) developed/implemented variants of a candidate reference measurement procedure (cRMP) based on isotope dilution-liquid chromatography-mass spectrometry. The sole constraint implemented was calibration with a common thyroxine primary calibrator. The RMPs were externally validated and assessed for comparability in round-robin trials using common samples, i.e., 5 lyophilized and 33 frozen native sera. At the same time, the performance of the cRMLs organized in a network was assessed. For uniform external quality assessment, common performance specifications were agreed on.

RESULTS

All cRMLs performed the cRMPs with fulfillment of the predefined specifications: total and between-laboratory CVs < or =2.0% and 2.5%, respectively, and a systematic deviation < or =0.9%, estimated with a target assigned from the mean of means obtained by the cRMLs. The mean expanded uncertainty for value assignment to the native sera was 2.1%.

CONCLUSIONS

A network of cRMLs, with externally conformed competence to properly perform RMPs, has been established. Performance specifications were defined and will form the basis for admittance of new network members. A serum panel, successfully targeted during the validation process, is available for split-sample measurements with commercial routine measurement procedures. The model can now be used for other measurands for which traceability to the Systeme International d'Unites is needed.

摘要

背景

确保/证明体外诊断的计量溯源性需要有特定被测量的参考测量系统(RMS),并且行业要有与合格参考测量实验室(RML)合作的可能性。在此,我们报告一项欧洲项目的结果,该项目旨在研究开发血清总甲状腺素RMS的可行性。

方法

四个候选RML(cRML)基于同位素稀释-液相色谱-质谱法开发/实施了候选参考测量程序(cRMP)的变体。实施的唯一限制是使用共同的甲状腺素一级校准品进行校准。这些RMP在外部进行了验证,并通过使用共同样本(即5份冻干血清和33份冷冻天然血清)的循环试验评估了可比性。同时,对以网络形式组织的cRML的性能进行了评估。为了进行统一的外部质量评估,商定了共同的性能规范。

结果

所有cRML执行cRMP时均符合预定义规范:总变异系数(CV)和实验室间CV分别≤2.0%和2.5%,系统偏差≤0.9%,该偏差是根据cRML获得的均值的平均值指定的目标进行估计的。将值赋予天然血清的平均扩展不确定度为2.1%。

结论

已建立了一个cRML网络,其在外部具备正确执行RMP的能力。定义了性能规范,这些规范将成为接纳新网络成员的基础。在验证过程中成功靶向的血清样本组可用于与商业常规测量程序进行的分割样本测量。该模型现在可用于其他需要溯源至国际单位制的被测量。