MR unit, Department of Radiodiagnostics and Interventional Radiology, Institute for Clinical and Experimental Medicine, Vídenská 1958/9, 140 00, Prague, Czech Republic.
MAGMA. 2004 Dec;17(3-6):359-62. doi: 10.1007/s10334-004-0077-1. Epub 2004 Dec 1.
A new design of RF coil based on a quasi-transverse electromagnetic field is described. The coil was developed for the acquisition of MR images of the rat spinal cord at 4.7 T. Different materials for the construction of the coil were tested, and the best results were obtained with Teflon. The design of the microstrip coil enables the investigator to change the length of the coil in a longitudinal direction and yields a relatively high signal-to-noise ratio due to the restricted field of view. Low RF field penetration depth also helps in suppressing motion artifacts generated by, e.g., breathing or heartbeats.