K-band EPR dosimetry: small-field beam profile determination with miniature alanine dosimeter.

作者信息

Chen Felipe, Graeff Carlos F O, Baffa Oswaldo

机构信息

Departmento de Física e Matemática, FFCLRP-Universidade de São Paulo, Av. Bandeirantes, 3900, 14040-901, Ribeirão Preto-SP, Brazil.

出版信息

Appl Radiat Isot. 2005 Feb;62(2):267-71. doi: 10.1016/j.apradiso.2004.08.036.

Abstract

The use of small-size alanine dosimeters presents a challenge because the signal intensity is less than the spectrometer sensitivity. K-band (24 GHz) EPR spectrometer seems to be a good compromise between size and sensitivity of the sample. Miniature alanine pellets were evaluated for small-field radiation dosimetry. Dosimeters of DL-alanine/PVC with dimensions of 1.5 mm diameter and 2.5 mm length with 5 mg mass were developed. These dosimeters were irradiated with 10 MV X-rays in the dose range 0.05-60 Gy and the first harmonic (1h) spectra were recorded. Microwave power, frequency and amplitude of modulation were optimized to obtain the best signal-to-noise ratio (S/N). For beam profile determination, a group of 25 dosimeters were placed in an acrylic device with dimensions of (7.5 x 2.5 x 1)cm3 and irradiated with a (3 x 3)cm2 10 MV X-rays beam field size. The dose at the central region of the beam was 20 Gy at a depth of 2.2 cm (build up for acrylic). The acrylic device was oriented perpendicular to the beam axis and to the gantry rotation axis. For the purposes of comparison of the spatial resolution, the beam profile was also determined with a radiographic film and 2 mm aperture optical densitometer; in this case the dose was 1 cGy. The results showed a similar spatial resolution for both types of dosimeters. The dispersion in dose reading was larger for alanine in comparison with the film, but alanine dosimeters can be read faster and more directly than film over a wide dose range.

摘要

使用小尺寸丙氨酸剂量计存在挑战,因为信号强度低于光谱仪的灵敏度。K波段(24吉赫兹)电子顺磁共振光谱仪似乎是样品尺寸和灵敏度之间的良好折衷。对微型丙氨酸颗粒进行了小场辐射剂量测定评估。开发了直径1.5毫米、长度2.5毫米、质量5毫克的DL-丙氨酸/聚氯乙烯剂量计。这些剂量计用10兆伏X射线在0.05 - 60戈瑞的剂量范围内进行辐照,并记录了一次谐波(1h)光谱。对微波功率、频率和调制幅度进行了优化,以获得最佳信噪比(S/N)。为了确定射野轮廓,将一组25个剂量计放置在尺寸为(7.5×2.5×1)立方厘米的丙烯酸装置中,并用(3×3)平方厘米的10兆伏X射线射野进行辐照。在2.2厘米深度处(丙烯酸的建成区)射野中心区域的剂量为20戈瑞。丙烯酸装置垂直于射野轴和机架旋转轴定向。为了比较空间分辨率,还用射线照相胶片和2毫米孔径光密度计确定了射野轮廓;在这种情况下剂量为1厘戈瑞。结果表明两种类型的剂量计具有相似的空间分辨率。与胶片相比,丙氨酸剂量读数的离散度更大,但在较宽的剂量范围内,丙氨酸剂量计比胶片读取速度更快、更直接。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验