Normative findings of electrically evoked compound action potential measurements using the neural response telemetry of the Nucleus CI24M cochlear implant system.

作者信息

Cafarelli Dees D, Dillier N, Lai W K, von Wallenberg E, van Dijk B, Akdas F, Aksit M, Batman C, Beynon A, Burdo S, Chanal J-M, Collet L, Conway M, Coudert C, Craddock L, Cullington H, Deggouj N, Fraysse B, Grabel S, Kiefer J, Kiss J G, Lenarz T, Mair A, Maune S, Müller-Deile J, Piron J-P, Razza S, Tasche C, Thai-Van H, Toth F, Truy E, Uziel A, Smoorenburg G F

机构信息

Cochlear Europe Ltd., London, UK.

出版信息

Audiol Neurootol. 2005 Mar-Apr;10(2):105-16. doi: 10.1159/000083366. Epub 2005 Jan 12.

Abstract

One hundred and forty-seven adult recipients of the Nucleus 24 cochlear implant system, from 13 different European countries, were tested using neural response telemetry to measure the electrically evoked compound action potential (ECAP), according to a standardised postoperative measurement procedure. Recordings were obtained in 96% of these subjects with this standardised procedure. The group results are presented in terms of peak amplitude and latency, slope of the amplitude growth function and ECAP threshold. The effects of aetiological factors and the duration of deafness on the ECAP were also studied. While large intersubject variability and intrasubject variability (across electrodes) were found, results fell within a consistent pattern and a normative range of peak amplitudes and latencies was established. The aetiological factors had little effect on the ECAP characteristics. However, age affected ECAP amplitude and slope of the amplitude growth function significantly; i.e., the amplitude is higher in the lowest age category (15-30 years). Principal component analysis of the ECAP thresholds shows that the thresholds across 5 electrodes can be described by two factors accounting for 92% of the total variance. The two factors represent the overall level of the threshold profiles ('shift') and their slopes across the electrode array ('tilt'). Correlation between these two factors and the same factors describing the T- and C-levels appeared to be moderate, in the range of 0.5-0.6.

摘要

按照标准化的术后测量程序,对来自13个不同欧洲国家的147名接受Nucleus 24人工耳蜗植入系统的成年受者,使用神经反应遥测技术测量电诱发复合动作电位(ECAP)。采用该标准化程序,在96%的受试者中获得了记录。以峰值幅度、潜伏期、幅度增长函数的斜率和ECAP阈值来呈现组结果。还研究了病因因素和耳聋持续时间对ECAP的影响。虽然发现受试者间和受试者内(跨电极)存在较大变异性,但结果仍处于一致模式,并建立了峰值幅度和潜伏期的正常范围。病因因素对ECAP特征影响不大。然而,年龄对ECAP幅度和幅度增长函数的斜率有显著影响;即,在最低年龄组(15 - 30岁)中幅度较高。对ECAP阈值的主成分分析表明,5个电极的阈值可由两个因素描述,这两个因素占总方差的92%。这两个因素分别代表阈值分布的总体水平(“偏移”)及其在电极阵列上的斜率(“倾斜”)。这两个因素与描述T级和C级的相同因素之间的相关性似乎为中等,在0.5 - 0.6范围内。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验