Scanning laser polarimetry retinal nerve fiber layer thickness measurements after LASIK.

作者信息

Zangwill Linda M, Abunto Teresa, Bowd Christopher, Angeles Raymund, Schanzlin David J, Weinreb Robert N

机构信息

Hamilton Glaucoma Center and the Department of Ophthalmology, University of California, San Diego, California 92093-0946, USA.

出版信息

Ophthalmology. 2005 Feb;112(2):200-7. doi: 10.1016/j.ophtha.2004.08.019.

Abstract

PURPOSE

To compare retinal nerve fiber layer (RNFL) thickness measurements before and after LASIK.

DESIGN

Cohort study.

PARTICIPANTS

Twenty participants undergoing LASIK and 14 normal controls.

METHODS

Retinal nerve fiber layer thickness was measured before LASIK and approximately 3 months after surgery in one eye each of 20 patients using a scanning laser polarimeter (GDx Nerve Fiber Analyzer) with fixed corneal compensation (FCC), one with variable corneal compensation (GDx VCC), and optical coherence tomography (OCT). Fourteen normal controls also were tested at baseline and approximately 3 months later.

MAIN OUTCOME MEASURES

Retinal nerve fiber layer thicknesses measured with the GDx FCC, GDx VCC, and OCT.

RESULTS

At baseline, mean (95% confidence interval [CI]) RNFL thicknesses for the GDx FCC, GDx VCC, and OCT were 78.1 microm (72.2-83.9), 54.3 microm (52.7-56.0), and 96.8 microm (93.2-100.5), respectively. In both LASIK and control groups, there were no significant changes between baseline and follow-up examinations in GDx VCC and OCT RNFL thickness measurements globally or in the superior and inferior quadrants (mean change, <5 microm for each instrument). In the control group, there also was no significant change in GDx FCC measurements between baseline and follow-up. In LASIK patients, significant reductions were observed in GDx FCC RNFL measurements. Average absolute values of the mean (95% CI) change in thickness were 12.4 microm (7.7-17.2), 15.3 microm (9.6-20.9), and 12.9 microm (7.6-18.1) for GDx FCC RNFL measurements superiorly, inferiorly, and globally, respectively (all Ps < or = 0.001).

CONCLUSIONS

LASIK does not seem to change RNFL thickness. Reduction in GDx FCC RNFL thickness measurements after LASIK is a measurement artifact and is most likely due to erroneous compensation for corneal birefringence. With scanning laser polarimetry, it is mandatory to compensate individually for change in corneal birefringence after LASIK to ensure accurate RNFL assessment.

摘要

目的

比较准分子激光原位角膜磨镶术(LASIK)前后视网膜神经纤维层(RNFL)厚度的测量结果。

设计

队列研究。

参与者

20例行LASIK手术的患者和14名正常对照者。

方法

使用具有固定角膜补偿(FCC)的扫描激光偏振仪(GDx神经纤维分析仪)、具有可变角膜补偿的仪器(GDx VCC)以及光学相干断层扫描(OCT),在20例患者的每只眼睛LASIK术前及术后约3个月测量视网膜神经纤维层厚度。14名正常对照者也在基线时及大约3个月后接受检测。

主要观察指标

用GDx FCC、GDx VCC和OCT测量的视网膜神经纤维层厚度。

结果

基线时,GDx FCC、GDx VCC和OCT测量的平均(95%可信区间[CI])RNFL厚度分别为78.1微米(72.2 - 83.9)、54.3微米(52.7 - 56.0)和96.8微米(93.2 - 100.5)。在LASIK组和对照组中,GDx VCC和OCT测量的RNFL厚度在全球范围以及上下象限的基线和随访检查之间均无显著变化(每种仪器的平均变化<5微米)。在对照组中,GDx FCC测量值在基线和随访之间也无显著变化。在LASIK患者中,观察到GDx FCC测量的RNFL有显著降低。GDx FCC测量的RNFL厚度平均(95% CI)变化的平均绝对值在上、下和全球范围分别为12.4微米(7.7 - 17.2)、15.3微米(9.6 - 20.9)和12.9微米(7.6 - 18.1)(所有P≤0.001)。

结论

LASIK似乎不会改变RNFL厚度。LASIK术后GDx FCC测量的RNFL厚度降低是一种测量假象,很可能是由于对角膜双折射的错误补偿。使用扫描激光偏振仪时,必须对LASIK术后角膜双折射的变化进行个体补偿,以确保准确评估RNFL。