Suppr超能文献

封闭式气管内吸痰系统对机械通气性能的影响。

The impact of closed endotracheal suctioning systems on mechanical ventilator performance.

作者信息

El Masry Ashraf, Williams Purris F, Chipman Daniel W, Kratohvil Joseph P, Kacmarek Robert M

机构信息

Department of Anesthesia and Critical Care, Massachusetts General Hospital, Boston 02114, USA.

出版信息

Respir Care. 2005 Mar;50(3):345-53.

Abstract

BACKGROUND

Closed endotracheal suctioning during mechanical ventilation is increasingly used, but its impact on ventilator function has not been fully studied.

METHODS

We evaluated the impact of closed suctioning with 11 critical-care ventilators, during assisted ventilation in pressure-support mode, pressure-assist/control mode, volume-assist/control mode, and during continuous positive airway pressure, with 2 suctioning pressures (-120 mm Hg and approximately -200 mm Hg), and with 2 tidal volumes (450 mL and 900 mL). We continuously measured airway pressure, flow at the airway, and pressure distal to the catheter tip, before, during, and after a single 15-second period of continuous suctioning.

RESULTS

No ventilator malfunctioned as a result of the closed suctioning. During suctioning, end-expiratory pressure markedly decreased in all modes, and peak flow increased in all modes except volume-assist/control (p < 0.001). Respiratory rate increased during suctioning in pressure- and volume-assist/control (p < 0.001) but not during pressure support or continuous positive airway pressure. Gas delivery was most altered during volume-assist/control with the smaller tidal volume (p < 0.05) and least altered during pressure-assist/control with the larger tidal volume.

CONCLUSION

There are large differences between the ventilators evaluated (p < 0.001). Closed suctioning does not cause mechanical ventilator malfunction. Upon removal of the suction catheter, these ventilators resumed their pre-suctioning-procedure gas delivery within 2 breaths, and, during all the tested modes, all the ventilators maintained gas delivery. However, closed suctioning can decrease end-expiratory pressure during suctioning.

摘要

背景

机械通气期间的密闭式气管内吸痰应用日益广泛,但其对呼吸机功能的影响尚未得到充分研究。

方法

我们使用11台重症监护呼吸机,在压力支持模式、压力辅助/控制模式、容量辅助/控制模式下的辅助通气期间以及持续气道正压通气期间,以2种吸痰压力(-120 mmHg和约-200 mmHg)和2种潮气量(450 mL和900 mL)评估密闭式吸痰的影响。在单次持续15秒吸痰的前、中、后,我们持续测量气道压力、气道流量以及导管尖端远端的压力。

结果

未因密闭式吸痰导致呼吸机故障。吸痰期间,所有模式下呼气末压力均显著降低,除容量辅助/控制模式外,所有模式下峰值流量均增加(p < 0.001)。压力辅助/控制和容量辅助/控制模式下吸痰期间呼吸频率增加(p < 0.001),但压力支持或持续气道正压通气期间呼吸频率未增加。在较小潮气量的容量辅助/控制模式下气体输送变化最大(p < 0.05),在较大潮气量的压力辅助/控制模式下气体输送变化最小。

结论

所评估的呼吸机之间存在很大差异(p < 0.001)。密闭式吸痰不会导致机械呼吸机故障。移除吸痰导管后,这些呼吸机在2次呼吸内恢复吸痰前的气体输送,并且在所有测试模式下,所有呼吸机均维持气体输送。然而,密闭式吸痰可在吸痰期间降低呼气末压力。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验