Chen Q, Mou S, Hou X, Ni Z
Research Center for Eco-Environmental Sciences, the Chinese Academy of Sciences, Beijing, 100085.
Se Pu. 1997 May;15(3):240-2.
A new high performance ion exchange chromatographic method for separation and determination of three organogermanium compounds beta-carboxyethylgermanium sesquioxide (I), beta-(alpha-methyl) carboxyethylgermanium sesquioxide (II) and di-(beta-carboxyethyl) germanium hydroxide (III) has been developed. A Dionex DX-300 Ion Chromatograph equipped with a Dionex PED-II pulsed electrochemical detector (conductivity mode), a Dionex AMMS-1 anion micromembrane suppressor, and a Dionex ACI advanced computer interface coupled with AI-450 chromatographic software was employed. The separation was achieved by using a Dionex IonPac AS4A-SC column as analytical column, sodium tetraborate solution as eluent, and sulfuric acid solution as regenerant. For reducing run time, a gradient program was chosen. The detection limits (S/N = 3, expressed as germanium) for the three compounds were 0.038mg/L (I), 0.035mg/L (II) and 0.025mg/L (III), respectively. The method has been applied to the analysis of two tonic oral drinks, and the average recoveries for the three compounds ranged from 95%-101%. The results obtained were in agreement with those of hydride generation atomic fluorescence spectrometry (HG-AFS).
已开发出一种新型高效离子交换色谱法,用于分离和测定三种有机锗化合物,即β-羧乙基锗倍半氧化物(I)、β-(α-甲基)羧乙基锗倍半氧化物(II)和二(β-羧乙基)锗氢氧化物(III)。使用配备了Dionex PED-II脉冲电化学检测器(电导模式)、Dionex AMMS-1阴离子微膜抑制器以及与AI-450色谱软件联用的Dionex ACI高级计算机接口的Dionex DX-300离子色谱仪。采用Dionex IonPac AS4A-SC柱作为分析柱,硼酸钠溶液作为洗脱液,硫酸溶液作为再生剂实现分离。为缩短运行时间,选择了梯度程序。三种化合物的检测限(S/N = 3,以锗计)分别为0.038mg/L(I)、0.035mg/L(II)和0.025mg/L(III)。该方法已应用于两种滋补口服饮料的分析,三种化合物的平均回收率在95% - 101%之间。所得结果与氢化物发生原子荧光光谱法(HG-AFS)的结果一致。