Characteristic resistance curves of aortic valve substitutes facilitate individualized decision for a particular type.

作者信息

Kuehnel Ralf-U, Puchner Reinhard, Pohl Alexander, Wendt Max O, Hartrumpf Martin, Pohl Manfred, Albes Johannes M

机构信息

Department of Cardiovascular Surgery Heart Center Brandenburg, Ladeburger Strasse 17, 16321 Bernau, Germany.

出版信息

Eur J Cardiothorac Surg. 2005 Mar;27(3):450-5; discussion 455. doi: 10.1016/j.ejcts.2004.12.003. Epub 2005 Jan 13.

Abstract

OBJECTIVE

Biological valves exhibit characteristic curves (CC) regarding transvalvular gradient, resistance, and effective orifice area when correlated with a physiological cardiac output range (CO). The slope of the curve of transvalvular resistance over a typical CO range characterizes the clinical performance of the valve. These information may support an individualized decision towards the most adequate valve type.

METHODS

In an extracorporeal mock circuit two types of stented biological aortic valves (constructed pericardial valves, Edwards Perimount: EP; porcine cusp valves, Medtronic Mosaic: MM) of 21, 23, and 25 mm were investigated. Mean transvalvular gradient was measured over a range of 1.9 to 7.2l/min CO at a simulated heart rate of 70 beats/min. Transvalvular resistance was calculated and presented as characteristic curves in a log-log-plot against cardiac output.

RESULTS

EP valves of all sizes demonstrated low slopes (resistance range; slope: 21 mm: 53-79 dynes s cm(-5); 0.29; 23 mm: 44-56 dynes s cm(-5); 0.12; 25 mm: 38-45 dynes s cm(-5); 0.12) while MM valves exhibited steep slopes (resistance range; slope: 21 mm: 46-169 dynes s cm(-5); 0.97; 23 mm: 36- -146 dynes s cm(-5); 0.95; 25 mm: 27-64 dynes s cm(-5); 0.68).

CONCLUSIONS

While constructed pericardial valves demonstrate sufficient hemodynamic performance especially in the higher CO range porcine cusp valves exhibited minor resistance in the lower CO range. Patients who exercise regularly may therefore profit from a pericardial valve while patients with a small body surface area and little exercise who therefore remain in the lower CO range may be adequately treated with a porcine cusp valve.