Kuehnel R-U, Pohl A, Puchner R, Wendt M O, Hartrumpf M, Pohl M, Albes J M
Department of Cardiovascular Surgery, Heart Center Brandenburg, Bernau, Germany.
Thorac Cardiovasc Surg. 2006 Mar;54(2):85-90. doi: 10.1055/s-2005-872854.
In an increasingly senescent population stented biological valves have regained renewed popularity because of the absence of anticoagulation, while the stented design allows for safe and easier implantation. Constructed bovine pericardial valves as well as valves with porcine cusps are used, both of which exhibit good clinical results although degeneration still appears. While clinical hemodynamic studies did not show particular differences between both valves types, the opening and closure behavior of native cusps and artificially constructed pericardial leaflets is different. It is unclear whether these phenomena account for differences in load and stress which may influence onset and course of degeneration.
Edwards Perimount (EP) and Medtronic Mosaic (MM) heart valves with diameters of 21 mm, 23 mm, and 25 mm were investigated in a pulse duplicator. Movements of the valves were visualized with a high-speed camera (1000 frames/sec). Mean transvalvular gradient (mm Hg), dissipated power (mW), and power transfer by stretching (mW), mean orifice area (mm2), opening time (ms), and closure time (ms) were analyzed in a range of cardiac outputs from 1.4 l/min to 6.3 l/min and 70 beats per minute.
Closure times were generally longer than opening times for both valve types. Opening time of EP valves was longer than opening time of the MM valves of the same size (EP23: 31.2 +/- 2.5 ms; MM23: 12.7 +/- 0.1 ms). With respect to closure times, however, there were no marked differences between all valves (EP23: 69.3 +/- 2.0 ms; MM23: 63.2 +/- 6.3 ms). Smaller sized Perimount valves exhibited lower mean transvalvular gradients than Mosaic valves of the same size (EP23: 7.21 +/- 0.07 mm Hg; MM23: 10.5 +/- 0.15 mm Hg). In larger sizes these differences diminished. Power transfer to the valve's structures was significantly enhanced in EP valves (EP23: 134 +/- 1.3 mW; MM23: 64 +/- 0.9 mW).
While valves with constructed pericardium showed lower mean transvalvular gradients, particularly in the smaller sizes, this valve type exhibited alterations of movement performance in contrast to porcine valves. It can be speculated that constant power transfer to the valve's structures may result in an earlier degeneration because of the impact of the increased load and stress on the suspension apparatus of the constructed pericardial leaflets.
在人口老龄化日益加剧的情况下,带支架的生物瓣膜因无需抗凝而再度受到欢迎,同时带支架的设计使植入更安全、更容易。使用的有构建的牛心包瓣膜以及带有猪瓣膜尖的瓣膜,尽管仍会出现退变,但两者都有良好的临床效果。虽然临床血流动力学研究未显示这两种瓣膜类型之间有特别差异,但天然瓣膜尖和人工构建的心包瓣叶的开闭行为不同。尚不清楚这些现象是否导致了负荷和应力的差异,而这可能会影响退变的发生和进程。
对直径为21毫米、23毫米和25毫米的爱德华兹Perimount(EP)和美敦力Mosaic(MM)心脏瓣膜在脉动复制器中进行研究。用高速摄像机(1000帧/秒)观察瓣膜的运动。在1.4升/分钟至6.3升/分钟的心输出量范围以及每分钟70次心跳的情况下,分析平均跨瓣压差(毫米汞柱)、消散功率(毫瓦)、拉伸传递功率(毫瓦)、平均开口面积(平方毫米)、开启时间(毫秒)和关闭时间(毫秒)。
两种瓣膜类型的关闭时间通常都比开启时间长。相同尺寸的EP瓣膜的开启时间比MM瓣膜长(EP23:31.2±2.5毫秒;MM23:12.7±0.1毫秒)。然而,就关闭时间而言,所有瓣膜之间没有明显差异(EP23:69.3±2.0毫秒;MM23:63.2±6.3毫秒)。较小尺寸的Perimount瓣膜的平均跨瓣压差低于相同尺寸的Mosaic瓣膜(EP23:7.21±0.07毫米汞柱;MM23:10.5±0.15毫米汞柱)。在较大尺寸时,这些差异减小。EP瓣膜中传递到瓣膜结构的功率显著增强(EP23:134±1.3毫瓦;MM23:64±0.9毫瓦)。
虽然带有构建心包的瓣膜平均跨瓣压差较低,尤其是在较小尺寸时,但与猪瓣膜相比,这种瓣膜类型表现出运动性能的改变。可以推测,由于增加的负荷和应力对构建心包瓣叶的悬吊装置的影响,持续传递到瓣膜结构的功率可能导致更早的退变。